Yodify Product Library
Add the Water Vapor Isotope Analyzer (H2O, δ2H, δ17O, δ18O) to your store or catalog  
Book Your Demo and See How
or create your store

Water Vapor Isotope Analyzer (H2O, δ2H, δ17O, δ18O)

Brand: Los Gatos Research, Inc

Now you can accurately measure δ17O, δ18O and δ2H in ambient water vapor (to over 60,000 ppm) rapidly with high precision in a measurement time of less than 0.5 seconds. Our Water Vapor Isotope Analyzer requires only 180 watts of power and does not require any sample preparation or user intervention, which enables long-term studies of water-vapor isotopes with unprecedented performance. In addition, the  Water Vapor Isotope Analyzer provides simultaneous measurements of water vapor mole fraction [H2O] with ppm-level precision. For measurements of liquid water as well as water vapor in one instrument, please refer to LGR's new Isotopic Water Analyzer.

LGR's  Water Vapor Isotope Analyzer works seamlessly with LGR's Water Vapor Isotope Standard Source (WVISS), which provides a controllable flow of humid air with known isotope values for  δ17O, δ18O and δ2H over a continuous range of H2O mole fractions (not just at a couple of concentrations).  The  Water Vapor Isotope Standard Source automatically and comprehensively characterizes the operation of the Water Vapor Isotope Analyzer (WVIA) in real time and without user intervention. Together, the WVIA and WVISS provide reliable, real-time, dual-inlet, absolute measurements of δ17O, δ18O and δ2H in ambient water vapor - no other combination provides comparable accuracy, precision and dynamic range.  The rugged, field-portable packaging makes this instrument ideally suited for a myriad of hydrological, atmospheric science, medical, and industrial monitoring applications.
Features
  • Only instrument to report δ²H, δ¹⁷O, δ¹⁸O and H₂O simultaneously
  • Automated dual-inlet operation provides unsurpassed accuracy, precision, and ease of use
  • Highest precision, reliable measurements at 2 Hz
  • High-resolution absorption spectra provide data validation and detailed system diagnostics
  • Now available in LGR’s Ultraportable package
  • Accurate measurements over an extremely wide range of mixing ratios and isotopic ratios
  • Proven fast flow time response allows isotopic flux studies
  • Enhanced performance models provide measurements with negligible drift
  • Insensitive to other ambient air constituents like methane or carbon dioxide

1 - 4 of 4 Part Numbers
911-0004
911-0004
  • Package Style - Rackmount Package
  • Measurable Components - d²H, d¹8O
911-0034
911-0034
  • Package Style - Rackmount Package
  • Measurable Components - d²H, d¹7O, d¹8O
912-0004
912-0004
  • Package Style - Benchtop Package
  • Measurable Components - d²H, d¹8O
912-0034
912-0034
  • Package Style - Benchtop Package
  • Measurable Components - d²H, d¹7O, d¹8O
1 - 4 of 4 Part Numbers

Specifications

Communication

  • Ethernet
  • RS-232 (Digital)
  • USB

Humidity

  • Ambient: 0 to 100 % RH, Non-Condensing

Maximum Drift

  • H₂O: 0.1 %
  • δ¹⁷O & δ¹⁸O: 0.28‰
  • δ²H: 0.88‰

Measurement Range

  • 4,000 to 60,000 ppm, Non-Condensing

Measurement Rate

  • Up to 2 Hz

Media

  • Liquid Water
  • Water Vapor

Mounting

  • Benchtop
  • Rack-Mounted

Operating Temperature

  • 0° to 40° C (32° to 104° F)

Power Requirement

  • 10-30 Vdc, 115/230 Vac, 65 W
  • 115/230 Vac, 50/60 Hz, 200 W

Precision (1σ, 10 Sec / 100 Sec)

  • H₂O: 0.2 % / 0.07 %
  • δ¹⁷O & δ¹⁸O: 0.15‰ / 0.05‰
  • δ²H: 0.5‰ / 0.2‰

Sample Temperature

  • -30° to 50° C (-22° to 122° F)

Accuracy

  • Total measurement uncertainty < 0.1 % (with WVISS)

Dimensions

  • 11" x 38" x 22" (Benchtop package)
  • 14" x 19" x 24" (Rackmount package)
  • 18.5’ x 14" x 7" (Portable package)

Weight

  • 36 kg (Rackmount package)
  • 40 kg (Benchtop package)
  • 15 kg (Portable package)