Yodify Product Library
Instantly add the 96001 Clamp-On Tester and 50,000+ more product series to your store or catalog  
Book Your Demo and See How
or create your store

Current clamp-on probe accessories for digital power meters, power analyzers, and digital multimeters. 

AC 400 Arms (600 A Peak) / AC 1000 Arms (1400 A Peak)
Features
  • Wide dynamic range: 0 Arms to 400 Arms (AC); Max. 600 Apk (AC)
  • Wide measurement frequency range: 20 Hz to 20 kHz (±5%)
  • Basic accuracy: ±1.0% of rdg ±0.2 mV
  • Voltage output model: 10 mV/A
  • Since this model is a voltage output type, you can use it for an oscilloscope.

Specifications

Accuracy

  • ± 1 % of Reading ± 0.2 mV (40 to 1,000 Hz) at 23° C ± 5° C (73.4° F ± 9° F), Sine Wave Input] (Amplitude)
  • ± 1.5 % of Reading ± 0.4 mV (20 to 40 Hz) at 23° C ± 5° C (73.4° F ± 9° F), Sine Wave Input] (Amplitude)
  • Within ± 3° C (± 5.4° F) (40 to 1,000 Hz) at 23° C ± 5° C (73.4° F ± 9° F), Sine Wave Input] (Phase)

Approvals & Certifications

  • CE

Approvals & Certifications

  • EN 61010-2-032

Cable Length

  • Approximately 8.2 feet (2.5 metres) (with Banana Plug) - Output Cord Length

Diameter

  • Maximum

External Magnetic Field Effect

  • 2 mV (0.2 A) or Less at 400 A/m

Frequency

  • 20 to 20,000 Hz

Humidity

  • 80 % Relative Humidity, Max.

Load Impedance

  • 100 kiloohms Min. / 100 PF, Max.

Measuring Range

  • 0 to 400 A ACrms (600 A Peak)

Operating Temperature

  • 0° to 50° C (32° to 122° F)

Output

  • 0-4 Vacrms (10 mV/A)

Output Impedence

  • Approximately 30 ohms

Storage Temperature

  • -20° to 60° C (-4° to 140° F), Non-Condensing

Supply Voltage

  • 3.7 kVac for One Minute (Core & Case, Core & Output Terminals) - Withstanding Voltage

Temperature Coefficient

  • 0.05 % of Full Scale/°C in a Range of [0° to 18° C (0° to 64.4° F)] & [28° to 50° C (82.4° to 122° F)]
Instantly populate your store or catalog 
CREATE YOUR SITE
or book a demo with the Yodify Team
show more