Q45C4 4E Conductivity Transmitter
Brand: Analytical Technology IncQ45C4 4E Conductivity Water Quality Transmitter is the solution for almost any water-based process monitoring.
ATi’s 4-Electrode Conductivity Transmitter is the answer for water quality monitoring almost any water-based process.
Drinking water, plating bath solutions, cooling water, process wash water, or virtually any other aqueous system can be monitored accurately and reliably.
The unique drive/control scheme used in the 4-Electrode System allows a single sensor configuration to be used in conductivity ranges from 0-200 μS to as high as 0-2,000 mS (0-2 S).
ATi’s Q45C4 4-Electrode Conductivity Transmitter’s “auto-ranging” feature enables the monitor to display the actual conductivity value during “overshoot” conditions.
For chemical mixing applications, a concentration display can be selected for a limited number of acids and bases.
If your application involves very low conductivity/resistivity measurement, ATi can provide the Q45C2 system that is used in process waters with conductivities below 20 μS.
Features- Measurement from 0-200 uS to 0-2.000 S.
- Loop-powered or battery-operated
- External 5-17 VDC version for solar applications
- 4-Electrode sensor design for fouling compensation
- Titanium electrodes for corrosion resistance
- Submersion, In-line, or Sanitary sensors
- Large, easy-to-read LCD display
- Clear, menu-driven user interface
- NEMA 4X (IP-66) enclosure
- Wall or pipe mounting bracket standard
- Built-in PID controller
Specifications
Display Range
- 0-200 µS
- 0-2.000, 0-20.00, 0-200.0, 0-2000 mS
- 0-2.000 S
Accuracy
- +0.5 % of Selected Range
Repeatability
- 0.2 % of Selected Range
Sensitivity
- 0.05 % of Selected Range
Linearity
- 0.1 % of Selected Range
Temperature Drift
- 0.1 % of Span/°C
Sensor Type
- Q25C4 4-Electrode
Temperature Input
- Pt1000 in Sensor for Temperature Compensation
Power
- 2-Wire Unit: 17-30 Vdc, 25 mA Max.
- Battery Unit: Two AA alkaline Batteries
- External DC: 5-17 Vdc, 10 mA Max.
Analog Outputs
- 2-Wire Unit: Single 4-20 mA DC, 450 ohms Max.
- Battery Unit: Two 0 - 2.5 Vdc, 50K ohms min.
- External DC: Two 0-2.5 Vdc, 50K ohms Min.
Output Isolation
- 600 V Galvanic Isolation
Enclosure
- NEMA 4X (IP-66) Polycarbonate, V-0 Flammability
Mounting Options
- Wall or Pipe Mounting Bracket Supplied
Conduit Entries
- Two PG-9 Cord Grips, 0.16 - 0.31" Diameter Cable
Ambient Temperature
- Storage: -40° to + 70° C
- Operating: -40° to + 60° C
Ambient Humidity
- 0-95 %, Non-Condensing
Size
- 4.4" H x 4.4" W x 3.5" D (111 x 111 x 90 mm)
Weight
- Battery Unit: 2 lb (0.9 Kg)
- 2-Wire or External DC unit: 1 lb (0.45 Kg)
Certifications
- CE marked to BS EN 61326-1:2006 Standard
- UL and CSA General Purpose
- Cooling Towers
- Plating Baths
- Potable Water
- Streams & Water
- Chemical Processing
- Fish Farms
- Wastewater
- Salt Water Intrusion