Brand: Status InstrumentsThe DM4500U accepts various types of sensors including Load Cell, Pt100, thermocouple, potentiometer and process current or voltage enabling the DM4500U to be used in a wide variety of applications.
Output options are provided that include NPN / PNP, 2 or 4 relays, current / voltage and RS485 comms. The unit can be fully programmed from the front panel buttons or via the RS485 comms option.
All parameters can be entered by pressing combinations of the three sealed front panel keys through a series of menus in helpful mnemonics or via the optional communications port.
The display can be programmed to read in three different colours and two levels of brightness. Red, Green and Orange. Colours can be programmed in a variety of different ways. For example alarm, normal running or high scale can be displayed in different colours.
Features- Suitable for load cell, temperature and process signals
- Internal power supply for loop and load cell excitation
- Tare functions for load cell
- Multi colour display
- Relay, NPN/ PNP, current / voltage output options
- RS485 comms option
Specifications
Brand
Accuracy
- ± 0.2° C (± 0.36° F) - Cold Junction
- 0.1 % of Full Scale ± 1-bit - mA/V Option
- 0.2 % of Reading at ± 0.6° C (± 1° F)
- 0.2 % of Reading at ± 1° C (± 2° F)
- 0.4 % of Reading at ± 0.6° C (± 1° F)
- 0.4 % of Reading at ± 1° C (± 2° F)
Approvals & Certifications
- UL 94 V-0
Communication
- RS-485
Configuration
- Asymmetric Differential
Current
- 100 µA, Max.
- 5 A, Max.
- 50 mA - Max.
- 8 A, Max.
Display
- -19999 to 19999, 5-Digits Tricolour 14 mm
- LED Display 4 for Functions & 4 for Outputs
Enclosure, Body Material
- Polycarbonate (UL 94 V-0)
Environmental Protection
- IP65
Excitation Voltage
- 10 V at 60 mA
- 24 V at 60 mA
- 5 V at 60 mA
Frequency
- 0.05 to 4 Hz (Cut-Off Frequency), 20 dB/Decade (Slope)
Function
- Tare for Load Cell
Humidity
- < 95 % Relative Humidity, Non-Condensing
Input
- < 1 mA DC - Pt100 Sensor Excitation
- ± 10 Vdc - Potentiometer Input, Process Voltage
- ± 15 mV - Load Cell Input
- ± 150 mV - Load Cell Input
- ± 20 mA DC - Process Current
- ± 30 mV - Load Cell Input
Input Impedance
- 1 Megaohm - Process Input for Voltage, Potentiometer Input
- 100 megaohms - Load Cell Input
- 15 ohms - Process Input for Current
Load
- ≤ 500 ohms, Max.
- ≥ 10 kiloohms, Max.
Operating Temperature
- -10° to 60° C (14° to 140° F)
Output
- 0-10 V
- 4-20 mA
Power
- 1,250 VA, Max.
- 150 W, Max.
- 192 W, Max.
- 2,000 VA, Max.
Power Supply
- 10.5-70 Vdc for DM4500/S2
- 100-300 Vdc for DM4500 / S1
- 22-53 Vac for DM4500/S2
- 85-265 Vac for DM4500/S1
Resistance
- 200 ohms - Potentiometer Min. Value
- 3 Milliohms, Max.
- 30 Milliohms, Max.
- 40 ohms/Cable - Max.
Resolution
- ± 15-bit - Conversion
- 0.001 % - Display Resolution
- 1 µA, Max.
- 1 µV, Max.
- 1 mV, Max.
- 13-bits for mA/V Option
Response Time
- 50 milliseconds
Sensor, Probe Type
- J Thermocouple
- K Thermocouple
- Pt100
- T Thermocouple
Storage Temperature
- -25° to 85° C (-13° to 185° F)
Supply Voltage
- 115-230 Vac, 0.2 A / 250 V - External Fuses (DIN 41661) for DM4500 / S1
- 125 Vdc, Max.
- 150 Vdc, Max.
- 24-48 Vac, 2 A/250 V - External Fuses (DIN 41661) for DM4500/S2
- 250 Vac, Max.
- 277 Vac, Max.
- 50 Vdc, Max.
Switch Type
- Relay
- SPDT
- SPST
- Transistor (NPN/PNP)
Temperature Coefficient
- 100 ppm/°C
Temperature Range
- -10° to 60° C (14° to 140° F) - Cold Junction Compensation
- -150° to 1,200° C (-238° to 2,192° F) for Thermocouples
Thermal Drift
- ± 0.05° C/°C
- 0.2 mV/°C - OPT4500/V
- 0.5 µA/°C - OPT4500/mA
Transfer Time
- 1 millisecond, Max.
- 10 milliseconds, Max.
Update Time
- 15 minutes