Yodify Product Library
Add the Ammonia Analyzer - Trace (NH3) to your store or catalog
Book Your Demo and See How
or create your store
Ammonia Analyzer - Trace (NH3)
Brand: Los Gatos Research, IncLGR's new Ammonia Analyzer - Trace (AAT) provides fast measurements of ammonia in ambient air with unparalleled precision, sensitivity and time response. No longer do you have to wait for minutes before getting a reliable reading – LGR’s AAT provides measurements at 10 Hz with sub-ppb precision. In addition, the AAT can report measurements quickly over a wide range of ammonia concentrations. For your convenience, we offer two versions of the AAT. If you require the fastest response rate possible (data rate up to 10 Hz), you will want to use our Ammonia Analyzer - Trace (Fast Flow). For applications requiring measurements of NH3 with highest accuracy, you’ll want the Ammonia Analyzer – Trace (Enhanced Performance).
The analyzer uses LGR’s patented Off-axis ICOS technology, a fourth-generation cavity enhanced absorption technique which employs a high-finesse optical cavity as the measurement cell. LGR’s technology has many proven advantages over conventional first-generation cavity ringdown spectroscopy (CRDS) techniques. In brief, since the laser beam does not have to be resonantly coupled to the measurement cell (i.e. precise beam alignment is not critical), LGR Analyzers are relatively inexpensive, simple to build, and inherently robust thermally and mechanically. In addition, since LGR’s technology can record reliable absorption spectra over a far wider range of optical depths (absorbance values) compared with CRDS, LGR Analyzers provide measurements over a much wider range of mole fractions (gas concentrations). Finally, due to the ease and simplicity of recording and quantifying the measured spectra using LGR’s Off-Axis ICOS, LGR Analyzers display the entire absorption spectra to the user in real time allowing for immediate system diagnostics and performance validation.
Features- Capability for eddy flux measurements (5 Hz flow time response, with external pump): allows observation of transient, time varying flows
- 0.2-ppb precision in 1 second
- Accurate measurements over a wide range of concentrations
- High resolution absorption spectra viewable for diagnostics
- Reports ammonia and water vapor simultaneously
- Reports water vapor up to 60,000 ppm
- Low power: ideal for field work
- New Enhanced Performance model provides lowest drift and highest accuracy
Specifications
Brand
Humidity
- Ambient: 0 to 100 % RH, Non-Condensing
Maximum Drift
- 0.2 ppb (Enhanced performance model)
Measurement Range
- 0.5 ppb to 10 ppm
Mounting
- Rack-Mounted
Operating Temperature
- 0° to 45° C (32° to 113° F) - EP Model
- 5° to 45° C (41° to 113° F) - Standard Model
Operational Range
- 0 to 100 ppm
Outputs
- Ethernet
- RS-232 (Digital)
- USB
Power Requirement
- 115/230 Vac, 50/60 Hz
Precision (1σ, 0.1 Sec / 1 Sec / 10 Sec)
- 0.7 ppb / 0.2 ppb / 0.08 ppb
Sample Temperature
- 0° to 50° C (32° to 122° F)
Power Consumption
- 300 W (EP model, steady State)
Dimensions
- Rackmount Package (Enhanced Performance model): 14" x 45" x 17"
Weight
- 58 kg (Standard model)
- 68 kg (EP model)