Yodify Product Library
Instantly add the 9091C Dissolved Hydrogen Analyzer and 50,000+ more product series to your store or catalog
Book Your Demo and See How
or create your store
9091C Dissolved Hydrogen Analyzer
Brand: Waltron Bull & Roberts, LLCThe Waltron 9091C Dissolved Hydrogen Analyzer is the industry standard for continuous measurement of dissolved hydrogen in water.
Measurement is microprocessor based and uses a potentiostatic three electrode system.
Important application points include primary loop of nuclear power plants, boiler feed-water in thermal power plants, denitrification of drinking water, and catalytic reduction of oxygen on noble metal surfaces in water treatment plants.
The dissolved hydrogen concentrations within the water/steam loop of a power plant gives quantitative and qualitative evidence of the rate of corrosion, corresponding to the buildup or destruction of the protective oxide layer inside pipes and boiler tubes.
Features- Portable system
- Fully automatic remote calibration
- Wide range analysis: 0 to 1,000 μg/L, 0 to 10 mg/L, 0 to 100 cc/kg
- Automatic temperature and flow compensation
- Fast response time (t90 = 30 sec)
- No zero point adjustment required
- Withstands pressures up to 8 bar (116 psi)
- Analog and digital outputs
- Minimal maintenance required
- Automatic operation
Specifications
Accuracy
- ± 3 % of Reading
Ambient Temperature Range
- 0° to 55° C (32° to 131° F)
Calibration
- Automatic Calibration
- Manual
Communication
- Serial Interface RS-232
Conductivity
- Sample Conductivity: ≥ 2 μS/cm, Salt Cell is Required if Conductivity is Less
Environmental Protection
- IP54
Flow Rate
- 3 to 18 LPH
Output
- 0-20 mA
- 4-20 mA
Power
- 100-240 Vac, 50/60 Hz
Pressure Range
- psig: 5 to 116 (kPa: 30 to 800, bar: 0.3 to 8)
Range
- 0 to 1,000 μg/L
- 0 to 10 mg/L
- 0 to 100 cc/kg
Response Time
- t90 = 30 seconds
Temperature Range
- Sample Temperature: 0° to 60° C (32° to 140° F)