PERMATRAN-C 4/30 Analyzer
Brand: MOCONThe PERMATRAN-C 4/30 CO2 permeation analyzer can test the Carbon Dioxide Transmission Rate (CO2TR) on both films and packages from low to high barriers.
It features two independent cells with interchangeable horizontal test cartridges that allow for a wide variety of sample sizes and packaging materials to be easily and accurately tested.
High Accuracy
- MOCON’s patented modulated infrared sensor is the only modulated sensor system on the market meeting ASTM F2476
- Low CO2TR detection levels down to 0.5 cc/(m2 x day)
- Fully automated temperature, flow controls simplify testing and provide unparalleled repeatability
High Throughput
- Modulated IR sensors measure CO2 gas concentrations in real time greatly reducing overall CO2TR test times compared to Zahm-Nagel CO2 testing methods
- Excellent correlation to Zahm-Nagel shelf life studies allowing equivalency while significantly reducing long term testing
- Automated test methods can start a test by a single press of a button
- TruSeal gas flushed perimeter sealing can eliminate the need for absolute zero in many tests, thus shortening overall test times
The PERMATRAN-C 4/30 analyzers can test the Carbon Dioxide Transmission Rate (CO2TR) from low to high barriers on films, packages and bottles.
Unlike manometric or Zahm-Nagel methods that require long periods of time to provide results the PERMATRAN-C 4/30 utilizes MOCON’s patented modulated IR sensor system allowing it to provide accurate results in much less time.
Designed for a wide range of barriers, this analyzer provides highly repeatable test results from 0.5 to 8,000 cc/ (m2 day).
The upper detection limit can be extended with the use of a reduced area film test cartridge (factor x9), or by using various sizes of masking.
The analyzer has industry leading features such as removable and interchangeable horizontal test cartridges, easy to use touch screen operator interface, and fully automated testing providing higher throughput in the lab with less training.
For more than 25 years, MOCON’s PERMATRAN series of analyzers have been the most trusted and longest lasting on the market.
The 4/30 delivers fast CO2TR results in less time than traditional methods helping to speed up your research and development of new materials or give you efficient QC verification of existing products.
The PERMATRAN-C 4/30 permeation analyzer is the perfect solution for rapid accurate CO2TR testing on low to high-level barriers.
Features- Removable Cartridges
- Easy sample preparation on the benchtop
- More consistent film loading into the analyzer
- One button release and loading
- Pneumatic clamping provides more repeatable results
- Interchangeable cartridge options that expand the testing capability
- TruSeal®
- Eliminates Individual Zero
- Reduces testing time (increases throughput)
- More repeatable results
- Automated Control Features
- Flow, Temp control
- Test Process Development and Recommendations
- Programmable Test Sequencing
- Test report printing and export
- Sensor Protection
- Touchscreen Interface
- Intuitive automated operation requires less training
- Built in multi-level user management and data security
- Event log & trouble shooting guide
- Save on valuable lab space by eliminating need for PC
- Modulated IR Sensor
- Short exam times reduces overall test time to increase throughput
- Long sensor life
- Accurate and repeatable results
- Most trusted modulated IR sensor that fully complies with ASTM F2476
- Warranty
- Full 100% parts and workmanship warranty for 12 months
- Exclusive 4-year warranty on the Modulated IR sensor
- Carbon Dioxide Permeation Analyzer
- Permeability Test Film and packages
- Complies with ASTM F2476
- 2 Cell Cartridge System
Specifications
Brand
Test Ranges (50 cm²)
- 0.5 to 8,000 cc/(m² x day)
- 0.03 to 516 cc/(100 in² x day)
Test Ranges (5.6 cm²)
- 4.5 to 71,000 cc/(m² x day)
- 0.29 to 4600 cc/(100 in² x day)
Test Ranges (5.0 cm² Masked)
- 5 to 80,000 cc/(m² x day)
Resolution (50 cm²)
- 0.0001 cc/(m² x day)
Test Range - Package
- 0.0025 to 40 cc/(pkg x day)
Repeatability
- 0.25 or 2.0 % cc/(m² x day), whichever is Larger
Temperature Control Range
- 10° to 40° C
Humidity (RH) Control Range Film Cell, Test Gas
- ± 0.2 % RH, Dry Only
Film Size
- 4” x 4” Standard, 2” x 2” Optional
Film Thickness
- Up to 120 Mil
Film Test Area
- 50 cm² Standard Cartridge
- 5.6 cm² Cartridge - Optional
Dimensions & Weight
- 15.5” H x 12” W x 23” D (39.4 cm x 30.4 cm x 58.0 cm)
- 95 lbs (43.1 kg)
Electrical Requirements
- Voltage: 100-240 Vac 50/60 Hz
- Maximum Power Draw: 700 VA
Laboratory Requirements
- Temperature - Operation: 22° C ± 2° C | Storage: 10° C to 30° C
- Humidity - Operation: 20 % to 80 % RH (Non-Condensing) | Storage: 5 % to 85 % RH (Non-Condensing)
Gas Supply Requirements
- Carrier Gas: Gas Composition - Nitrogen (99.7 % N₂ or better)
- Test Gas: Gas Composition - Carbon Dioxide (99.7 % or better)
Connectivity
- Remote Access and Control (requires Optional Perm-Net Light Software)
- Standard Ethernet Port (100T)
- Dual USB 2.0 Ports (for Connection of Keyboard, Mouse or Memory Devices)
Compliance Standards
- This product is UL and CSA listed using the ETL mark from Intertek Group plc. and is certified to conform to CE
- Conforms to the following internationally recognized safety standards: IEC 61010-1:2010 Third Edition | UL 61010-1:2012 Third Edition | CAN/CSA C22.2 No. 61010-1:2012 Third Edition
- Conforms to the following internationally recognized EMC standards: EN 61326-1:2013 | EN 61000-3-2:2006, A1:2009, A2:2009 | EN 61000-3-3:2013
- Conforms to EU Directive 2011/65/EU RoHS Compliance.
Cartridges & Options
- Reduced Area Cartridge
- Blister Cartridge
- Package Testing Adaptor
- PackRack Package Test Fixture
- Closure Test Cell
- 1 Liter Bottle Capture Volume Test Cell
- 2 Liter Bottle Capture Volume Test Cell
Accessories
- Consumables: Reference Films, Masks
Software
- PermWare Data Collection, Data Management and Instrument Status Dashboard
- Remote Viewing Software from a PC - PermNet Lite
- 21 CFR Part 11 User and Data Security
Service & Support Options
- IQ/OQ Documentation & Validation
- Basic Preventative Maintenance Plan
- Plus Preventative Maintenance Plan
Test Standard Compliance
- ASTM F2476
- DIN 53380-4
- ISO 15105-2
- Test Methodology: Following ASTM F2476-13
Test Methodology
- Standard Test Method for the Determination of Carbon Dioxide Gas Transmission Rate (CO2TR) Through Barrier Materials Using an Infrared Detector
Summary of Test Method
- The carbon dioxide gas transmission rate is determined after the sample has equilibrated in a dry-test environment. In this context, a “dry” environment is considered to be one in which the relative humidity is less than 1 %.
- The specimen is mounted as a sealed semi-barrier between two chambers at ambient atmospheric pressure. One chamber is slowly purged by a stream of nitrogen and the other chamber with carbon dioxide. As carbon dioxide gas permeates through the film into the nitrogen carrier gas, it is transported to an infrared detector where an electrical output is produced whose magnitude is proportional to the amount of CO2 flowing into the detector per unit of time.
- Packaging Films
- Monolayer
- Multi-layer
- Coated Films
- Packages
- Common Food Packages
- Bottles/Closures
Documentation
Brochures
Manuals
- A Comparison of the ASTM F1115 and F2476 Test Methods for CO₂ TR Measurement pdf 283 KB
- AN-PPS-1018 Critical Bag Material CO₂TR Specification in Medical Procedures pdf 346 KB
- How to Test Bottle CO₂ Transmission Rate for Shelf-Life Estimation pdf 164 KB
- MOCON AMECare Training & Webinars pdf 192 KB
- MOCON Permeation Accessory Sheet pdf 254 KB
- MOCON Value Promise & Service Guarantee pdf 135 KB