The M&C liquid-stop LB-1... is suitable for protecting analyzers against inrush of liquids out of the gas conditioning unit which is mounted before the analyzer. 

This avoids major damages of the analyzer. The most useful position of the liquid stop LB-1.. is behind the gas conditioning unit, directly before the flowmeter of the analyzer.
Features
  • Sure protection against liquid inrush
  • Reliable removal of liquid coming through
  • Also suitable for high pressures
  • Available in stainless steel and PVDF
  • Easy change of the hydrophobic protective membrane
  • With wall mounting holder

1 - 2 of 2 Part Numbers
03 F 4000
03 F 4000
  • Version - LB-1SS
03 F 4005
03 F 4005
  • Version - LB-1PV
1 - 2 of 2 Part Numbers

Specifications

Ambient Temperature Range

  • 0° to 60° C (32° to 140° F)

Differential Pressure

  • Differential Pressure with Clean Protection Membrane, Medium Air, 20° C (68° F) - psi: 0.725 to 1.45 (kPa: 5 to 10, Mbar: 50 to 100)

Flow Rate

  • Differential Pressure with Clean Protection Membrane, Medium Air, 20° C (68° F): 50 to 100 NL/hr
  • Gas Flow: Max.

Materials of Construction

  • Material of Gas Bearing Parts: 316Ti Stainless Steel, 1.4542 - Version LB-1SS
  • Material of Gas Bearing Parts: FPM
  • Material of Gas Bearing Parts: Glass
  • Material of Gas Bearing Parts: Polyester
  • Material of Gas Bearing Parts: PTFE
  • Material of Gas Bearing Parts: PVDF - Version LB-1PV

Mounting

  • Wall Mounting

NPT Connection

  • Sample Gas/Drain Connections: 1/8”i Female NPT DIN ISO 228/1

O-Ring Material

  • FPM

Pressure Range

  • Gas Pressure - psi a: 4.351 to 145 (kPa: 30 to 1,000, bar: 0.3 to 10) ∆P, Max.

Ring Material

  • Flat Ring: Teflon

Storage Temperature

  • -25° to 80° C (-13° to 176° F)