Yodify Product Library
Instantly add the LB-1PV/LB-1SS Liquid Stop and 50,000+ more product series to your store or catalog
Book Your Demo and See How
or create your store
LB-1PV/LB-1SS Liquid Stop
Brand: M&C TechGroup Germany GmbHThe M&C liquid-stop LB-1... is suitable for protecting analyzers against inrush of liquids out of the gas conditioning unit which is mounted before the analyzer.
This avoids major damages of the analyzer. The most useful position of the liquid stop LB-1.. is behind the gas conditioning unit, directly before the flowmeter of the analyzer.
Features- Sure protection against liquid inrush
- Reliable removal of liquid coming through
- Also suitable for high pressures
- Available in stainless steel and PVDF
- Easy change of the hydrophobic protective membrane
- With wall mounting holder
Specifications
Ambient Temperature Range
- 0° to 60° C (32° to 140° F)
Differential Pressure
- Differential Pressure with Clean Protection Membrane, Medium Air, 20° C (68° F) - psi: 0.725 to 1.45 (kPa: 5 to 10, Mbar: 50 to 100)
Flow Rate
- Differential Pressure with Clean Protection Membrane, Medium Air, 20° C (68° F): 50 to 100 NL/hr
- Gas Flow: Max.
Materials of Construction
- Material of Gas Bearing Parts: 316Ti Stainless Steel, 1.4542 - Version LB-1SS
- Material of Gas Bearing Parts: FPM
- Material of Gas Bearing Parts: Glass
- Material of Gas Bearing Parts: Polyester
- Material of Gas Bearing Parts: PTFE
- Material of Gas Bearing Parts: PVDF - Version LB-1PV
Mounting
- Wall Mounting
NPT Connection
- Sample Gas/Drain Connections: 1/8”i Female NPT DIN ISO 228/1
O-Ring Material
- FPM
Pressure Range
- Gas Pressure - psi a: 4.351 to 145 (kPa: 30 to 1,000, bar: 0.3 to 10) ∆P, Max.
Ring Material
- Flat Ring: Teflon
Storage Temperature
- -25° to 80° C (-13° to 176° F)