Yodify Product Library
Instantly add the Series 631B Capsuhelic® Wet/Wet Differential Pressure Transmitter and 50,000+ more product series to your store or catalog
Book Your Demo and See How
or create your store
Series 631B Capsuhelic® Wet/Wet Differential Pressure Transmitter
Brand: Dwyer Instruments, LLCThe Dwyer SERIES 631B Capsuhelic® Differential Pressure Transmitter monitors differential pressure of air and compatible gases and liquids with accuracy.
The design employs converting pressure changes into a standard 4-20 mA output signal for two wire circuits. Digital push-button, zero and span adjustments are easily accessed on the front cover.
The Series 631 Differential Pressure Transmitter is designed to meet NEMA 4X (IP66) construction. Robust housing offers 500 psi static pressure rating on ranges down to 0.5 inches WC.
Features- Gauge capsule permits high-pressure usage with small differentials
- Versatile device for liquid or gas supports designs requiring more precise measurements in support of application
- Zero and range adjustments outside of gauge means no disassembly in normal service
- NEMA 4X rated enclosure provides protection in harsh environments permitting outdoor monitoring or in areas where dust and particulate matter exists
1 - 10 of 10 Part Numbers
631B-0
- Pressure Range - Inches WC: 0 to 0.5
- Certificates - None
631B-0-NIST
- Pressure Range - Inches WC: 0 to 0.5
- Certificates - NIST Traceable Calibration Certificate
631B-1
- Pressure Range - Inches WC: 0 to 1
- Certificates - None
631B-1-NIST
- Pressure Range - Inches WC: 0 to 1
- Certificates - NIST Traceable Calibration Certificate
631B-2
- Pressure Range - Inches WC: 0 to 2
- Certificates - None
631B-2-NIST
- Pressure Range - Inches WC: 0 to 2
- Certificates - NIST Traceable Calibration Certificate
631B-3
- Pressure Range - Inches WC: 0 to 5
- Certificates - None
631B-3-NIST
- Pressure Range - Inches WC: 0 to 5
- Certificates - NIST Traceable Calibration Certificate
631B-5
- Pressure Range - Inches WC: 0 to 25
- Certificates - None
631B-5-NIST
- Pressure Range - Inches WC: 0 to 25
- Certificates - NIST Traceable Calibration Certificate
1 - 10 of 10 Part Numbers
Specifications
Brand
Accuracy
- Gauge: ± 3 % of Full Scale at 21.1° C (70° F)
- Transmitter Output: ± 2 % of Full Scale (Includes Linearity, Hysteresis & Repeatability)
Applications
- Energy/Water Management
- Liquid Level in Water Storage Tanks
- Refrigeration Equipment
Approvals & Certifications
- CE
Electrical Connectors
- M12 Circular 4-Pin Connector
Environmental Protection
- IP66
- NEMA 4X
Loop Resistance
- 0 to 1250 ohm
Media
- Compatible Gases
- Compatible Liquids
NACE & NIST
- NIST Traceable Calibration Certificate
NPT Connection
- 1/4” NPT Female
Output
- 4-20 mA
Power Requirement
- 10-35 Vdc
Pressure Range
- Inches WC: 0 to 25
Pressure Type
- Differential
Response Time
- 50 milliseconds
Sensor Orientation
- Diaphragm in Vertical Position
Stability
- ± 1 % of Full Scale Per Year
Temperature Effect
- Zero & Span: 0.025 %/°F (0.045 %/°C)
Temperature Range
- -6.67° to 48.9° C (20° to 120° F)
Wetted Materials
- 300 Stainless Steel
- Brass
- Silicone