Yodify Product Library
Add the Model 3010BR Trace Hydrogen Sulfide Analyzer to your store or catalog
Book Your Demo and See How
or create your store
Model 3010BR Trace Hydrogen Sulfide Analyzer
Brand: Advanced Micro Instruments, IncThe AMI Model 3010BR Trace Hydrogen Sulfide (H2S) Analyzer provides H2S monitoring in an unprecedented low cost, compact size, virtually maintenance-free package.
Using AMI’s patented cell block technology, it allows a specially modified electrochemical sensor to operate on oxygen-free sample gas, eliminating the need for hazardous-waste generating lead acetate tape.
The Model 3010BR provides a complete, industry proven electronics package and integrated sample system. It is offered for about a third the price of competitive models.
AMI’s patented cell block allows the analyzer to be built with almost no possibility of internal leaking, with minimal volume and with front-panel sensor access.
All sample handling components – the flow meter, needle valve and a four-way sample/span/off selection valve – are integrated into a solid metal block.
Connections between them are drilled passages in place of long lengths of tubing.
The result is a highly reliable sample system with all necessary components provided, and a fast response time – hydrogen sulfide is a “sticky” gas so minimizing the sample path is vital to get the least possible wetted surface area.
Features- High resolution 3-1/2 digit LCD
- 2 fully adjustable Hydrogen Sulfide (H2S) Concentration alarms
- Alarm delays
- Alarm hold off/Bypass
- RFI protected
- 1-5 Vdc and 4-20 mA. isolated Analog output signals
- USB virtual comport and Modbus Bidirectional RS485 communication for advanced features
Specifications
Approvals & Certifications
- CSA Class I, Div. 1, Groups B, C, D
- UL 913, UL 1203
Power Requirement
- 10-28 Vdc / 115 Vac < 70 mA @ 12 Vdc Non-Heated, < 24 W @ 12 Vdc with Heated Option
Pressure Range
- Inlet Gas Pressure - psig: 0.5 to 150 (kPa: 3 to 1,034, bar: 0.03 to 10.34)
Process Connection - Other
- 1/4" Stainless Steel Compression Fittings for Sample, Span Gas & Exhaust
Range
- 0 to 10 ppm
- 0 to 100 ppm
- 0 to 200 ppm
- 0 to 50 ppm
Repeatability
- ± 1 % of Range or ± 0.2 ppm of H₂S, whichever is greater