Yodify Product Library
Instantly add the PGC5000C Smart Oven Process Gas Chromatograph and 50,000+ more product series to your store or catalog
Book Your Demo and See How
or create your store
PGC5000C Smart Oven Process Gas Chromatograph
Brand: ABB GroupABB process gas chromatographs perform real-time analysis of the chemical composition of a process sample or stream -- crucial for on-spec processes and productivity.
PGC5000C Smart Ovens target complex applications requiring multiple detectors for maximum application densification.
The ABB PGC5000 Series process gas chromatograph is based on over 50 years of innovation in on-line process gas chromatographs. It simplifies gas chromatography and increases reliability. It provides everything you need to know about your GC – at your fingertips.
The PGC5000 Series has been designed for distributed control across multiple analytical Smart OvensTM for limitless application approaches.
There are two ovens, PGC5000B and PGC5000C, suitably equipped for the type of application engineered.
PGC5000C Smart OvenTM technology supports applications that require dual detectors with up to six analytical valves.
Advanced pressure, temperature and stream control software executes analytical methods required for analyses.
A single PGC5000A Controller can support up to two PGC5000C Smart OvensTM increasing applications flexibility and offering maximum application density.
Features- Appropriately sized (28% larger) and designed for maximum application densification
- Perfect for multiplexing multiple streams’ analyses in a single analytical oven
- Designed for complex application measurements
- Multiple detector configuration – sTCD, mTCD, FID and FPD capability
Specifications
Brand
Environmental Protection
- IP54 (NEMA 3 Equivalnet)
Ambient Temperature
- Oven Temperature: 30° to 180° C (Settings and display in °C only)
- 0° to 50° C (32° to 122° F)
Humidity
- 95 % Relative Humidity, Non-Condensing
Mounting
- Wall: 33 mm (1.3 in.) from wall with brackets
EMI / RFI
- Conform to ClassA industrial environment
Certifications & Approvals
- ATEX / IEC / CN / KO: Zone 1: CE 0344, II2G, Ex de py IIB+H₂ T4 - T2, Zone 2: CE, II3G Ex de nA nL IIB+H₂ T4 - T2 Ex de px IIB+H₂ T4 - T2 (Optional), with X-purge Power interlock
- CSA / NRTL: Class I, Div. 1, gas groups B, C, D with Type Y-purge, Class I, Div. 2, gas groups B, C, D Temperature code T4 - T2
Power Requirement
- 100-240 Vac, 50/60 Hz
Power Consumption
- 900 W steady-state Operation
- 1,600 W Startup
- Typical, varies with installed options
Tube Connection
- Supply: 3/8”, minimum
Pressure
- Inlet: 1034 kPa (150 psig) Max.
- Inlet: Set Point 69 kPa (10 psig) minimum
- Supply: 551.6 kPa (80 psig)
Flow Rate
- Steady State purge: 127 to 147 L/min (4.5 to 5.2 ft³/min) at 20° C, Y-purge Types
Material of Construction
- Tube: 316 Stainless Steel
- Oven liner: Stainless Steel
Resolution
- Setpoint: 1° C
Stability
- Ambient Range: ± 1.0° C
- Steady ambient: ± 0.1° C
Accuracy
- 2 % (0-100 psig)
Repeatability
- ± 0.05 psig
Media
- H₂, He, N₂, Air, Ar
- No Liquids, corrosives, combustibles, O₂
Documentation
Brochure
- Bulletin - PGC5000 Series - Factory notice - Safety reminder pdf 398 KB
- Bulletin - PGC5000 Series Version 4.2.1 Release is now available pdf 151 KB
- Bulletin - Time Synchronization in PGC5000 Data System pdf 218 KB
- Bulletin - VistaSTAR with PGC5000 Version 4 pdf 236 KB
- Course description - Z234e Process Gas Chromatography | External version pdf 49 KB
- Windows Server 2003 OS - Support ended on July 14, 2015 pdf 210 KB