Yodify Product Library
Instantly add the GM90PS Power Supply Module and 50,000+ more product series to your store or catalog
Book Your Demo and See How
or create your store
GM90PS Power Supply Module
Brand: Yokogawa Electric CorporationThe Modular GM90PS is a product to the Yokogawa, the SMARTDAC+ GM Data Acquisition System is a data logger that excels in versatility and expandability. The main unit includes on-board memory for data acquisition and also supports SD cards for external storage.
The system consists of a Data Acquisition Module (GM10), Power Supply Module (GM90PS), and Module Base (GM90MB), which is used to mount a variety of modules. SMARTDAC+ GM utilizes the same modules as the GX/GP series of SMARTDAC+.
Features- Scalability: Up to 420 channels per system Plug and lock modules
- Ease of Use: Web-based configuration Live web-based data viewing
- Mobile Connectivity: Bluetooth Mobile application
- Reliability: Secure data storage High accuracy measurement Automatic backfill function (GA10 Data Logging Software)
- Noise Tolerance: Electromagnetic relay module
Specifications
Ambient Temperature Range
- -20° to 60° C (-4° to 140° F)
Approvals & Certifications
- CE
- CSA 22.2 No. 61010-1, CSA 22.2 No. 61010-2-030-12
- EMC Directive: EN 61326-1 Class A Table 2, EN 61000-3-2, EN 61000-3-3, EN 55011 Class A Group 1
- EMC: EN 301 489-1, EN 301 489-17, EN 61326-1
- KC
- LVD: EN 61010-1, EN 61010-2-030
- RCM: AS/NZS4268, AS/NZS2772.2
- UL 61010-1, UL 61010-2-030 (CSA NRTL/C)
- WEEE Directive Support
Approvals & Certifications
- R & TTE Directive - EN 61010-1, EN 61010-2-030, EN 62311
Connection Style
- Screw Terminal
Frequency
- 50 Hz ± 2 % - Power Frequency
- 60 Hz ± 2 % - Power Frequency
Humidity
- 20 to 85 % Relative Humidity, Non-Condensing - Ambient Humidity
Insulation Resistance
- 20 megaohms or more at 500 Vdc between Power Terminal & Earth
Mounting Position
- Any Orientation
Operating Voltages
- 180-264 Vac - Operating Supply Voltage
- 90-132 Vac - Operating Supply Voltage
RFI/EMI Effect
- 400 A/m or Less (DC & 50/60 Hz) - Magnetic Field
Shock
- When Off, 500 m/sec² or Less, Approximately 10 milliseconds, 3 times in 6 Directions (± X, ± Y, ± Z)
- When On, 98 m/sec² or Less, 11 milliseconds, 3 times in 6 Directions (± X, ± Y, ± Z)
Supply Voltage
- 100-240 Vac - Rated Supply Voltage
- 3,000 Vac (50/60 Hz), 1 Minute - Withstand Voltage between Power Terminal & Earth
Update Time
- 30 minutes - Warm-Up Time
Vibration
- 5 ≤ F < 8.4 Hz Amplitude 3.5 mm Peak