8232 Chlorine Sensor
Brand: Burkert Fluid Control SystemsThe Type 8232 from Burkert is an electrochemical sensor designed for the measurement of the free chlorine concentration generated from inorganic sources (chlorine gas, sodium hypochlorite,...).
The Type 8232 offers a range of three models:
- The chlorine sensor “Trace” (zero-chlorine) with three electrodes, is suitable for the measuring of chlorine at very low concentrations. The membrane of this sensor is less affected by biofouling. The sensor works in water without chlorine up to four weeks. It is available with a voltage output on a four-pin fixed connector
- The chlorine sensor with three electrodes, has a sensor having a greatly reduced pH dependency. This model delivers a current output on a M12 connector and is designed to be used in a swimming pool, drinking or sea water. The fluid must contain a minimum chlorine concentration (≥ 0.1 ppm)
- The chlorine sensor with two electrodes, delivers a current output on a M12 connector. It is designed to be used in swimming pool, drinking or process water. The fluid to be measured must not contain any surfactants or abrasive particles and its pH value must be at a constant level. The fluid must contain a minimum chlorine concentration (≥ 0.1 ppm)
- Chlorine sensor “Trace” with three electrodes for measuring at low concentrations
- Chlorine sensor with three electrodes and greatly reduced pH dependency
- Chlorine sensor with two electrodes for standard applications at constant pH
1 - 3 of 3 Part Numbers
565 164
- Version - Chlorine Sensor ” Trace” (Zero-Chlorine)
- Measuring Range - 0.005 to 2 ppm
- Output - 0-2,000 mV (Max. 2,500 mV)
- Electrical Connection - 4-Pin Fixed Hermaphroditic Connector
568 523
- Version - Chlorine Sensor with 3 Electrodes
- Measuring Range - 0.01 to 20 ppm
- Output - 4-20 mA
- Electrical Connection - 5-Pin M12 Male Connector
568 524
- Version - Chlorine Sensor with 2 Electrodes
- Measuring Range - 0.01 to 20 ppm
- Output - 4-20 mA
- Electrical Connection - 5-Pin M12 Male Connector
1 - 3 of 3 Part Numbers
Specifications
Ambient Temperature Range
- Operation & Storage (Electrolyte): 5° to 35° C (41° to 95° F)
- Operation & Storage (Probe): 5° to 40° C (41° to 104° F)
Applications
- Monitoring Absence of Chlorine in Reverse Osmosis Systems - Chlorine Sensor ”Trace”
- Monitoring Chlorine (At Fluctuating PH - Chlorine Sensor with 3 Electrodes, at Constant PH - Chlorine Sensor with 2 Electrodes)
Approvals & Certifications
- CE
- EU Directives
Cap Material
- Membrane Cap: PVC
Consumption
- Current: 20 mA Approximately - Chlorine Sensor ”Trace”
- Current: 4 mA Approximately, Max.
Electrical Connectors
- 4-Pin Fixed Hermaphroditic Connector - Chlorine Sensor ”Trace”
- 5-Pin M12 Connector (Male) - Chlorine Sensor with 3 Electrodes, Chlorine Sensor with 2 Electrodes
- Screw Terminal
Flow Rate
- Fluid: 15 to 30 L/hr
Humidity
- < 90 % Relative Humidity, without Condensation
Loop Impedance
- 50 ohms at 12 Vdc, Max.
- 900 ohms at 30 Vdc, Max.
Materials of Construction
- 316Ti Stainless Steel, 1.4571 - Chlorine Sensor ”Trace”, Chlorine Sensor with 3 Electrodes
- Halide Reference Electrode: Silver
- Membrane Holder: Stainless Steel
- PVC - Chlorine Sensor ”Trace”
- PVC-U - Chlorine Sensor with 3 Electrodes, Chlorine Sensor with 2 Electrodes
- Working Electrode: Gold
Measuring Range
- 0.005 to 2 ppm - Chlorine Sensor ”Trace”
- 0.01 to 20 ppm - Chlorine Sensor with 3 Electrodes, Chlorine Sensor with 2 Electrodes
Media
- Water (6.5 to 9 PH - Chlorine Sensor ”Trace”, 4 to 9 PH - Chlorine Sensor with 3 Electrodes, 6 to 8 PH - Chlorine Sensor with 2 Electrodes)
O-Ring Material
- Sealing: NBR
Operating Voltages
- 12-30 Vdc, Filtered & Regulated - Chlorine Sensor with 3 Electrodes, Chlorine Sensor with 2 Electrodes
- 9-30 Vdc, Filtered & Regulated - Chlorine Sensor ”Trace”
Output
- 4-20 mA - Chlorine Sensor with 3 Electrodes, Chlorine Sensor with 2 Electrodes
- Analog: 0-2,000 mV, Max.
Pressure Range
- psi: 0 to 14.5 (kPa: 0 to 100, bar: 0 to 1) - Max.
- psi: 0 to 43.51 (kPa: 0 to 300, bar: 0 to 3) - Max.
- psi: 0 to 7.25 (kPa: 0 to 50, bar: 0 to 0.5) - Max.
Process Temperature
- 0° to 40° C (32° to 104° F) - Chlorine Sensor ”Trace”
- 0° to 45° C (32° to 113° F) - Chlorine Sensor with 3 Electrodes, Chlorine Sensor with 2 Electrodes
Resolution
- Sensor: 0.001 ppm - Chlorine Sensor ”Trace”
- Sensor: 0.01 ppm - Chlorine Sensor with 3 Electrodes, Chlorine Sensor with 2 Electrodes
Response Time
- t90: 120 seconds - Chlorine Sensor ”Trace”, Chlorine Sensor with 3 Electrodes
- t90: 30 seconds - Chlorine Sensor with 2 Electrodes
Ring Material
- Snap, Slide Ring: PETP
Shaft Material
- PVC
- Monitoring Chlorine (at Fluctuating PH - Chlorine Sensor with 3 Electrodes, at Constant PH - Chlorine Sensor with 2 Electrodes)
- Monitoring Absence of Chlorine in Reverse Osmosis Systems - Chlorine Sensor "Trace"