Yodify Product Library
Instantly add the Compact Density Meter and 50,000+ more product series to your store or catalog
Book Your Demo and See How
or create your store
Compact Density Meter
Brand: Emerson Electric Co.Providing precision density and temperature for custody transfer and concentration measurement, the Compact Density Meter (CDM) features simple system integration, health diagnostics and flow-rate indication to help verify sample integrity.
Features- Achieve superior real-world performance with elevated and combined temperature and pressure calibration.
- Ensure sample integrity with flow-rate indication.
- Reduce measurement errors and minimize negative installation effects with case RTDs that adjust for ambient temperature.
- Rely on compliance and compatibility to global standards - NACE, NORSOK, MID, ISO/IEC 17025.
- Quickly install and verify meter health with intelligent internal diagnostics.
- Support multiple simultaneous protocols for connection to DCS, PLC and flow computers.
- Expand usage with a hazardous-area-approved head-mounted transmitter that supports local configuration and display.
- Interact with our 3D model to learn more about our Compact Density Meters.
- Interact with our detailed schematic tool to learn more about our Dedicated Density and Viscosity Meters.
Specifications
Brand
Case Temperature Sensors
- Class B RTD (Quantity 3)
Density Accuracy
- ± 0.1 kg/m³ (± 0.0001 g/cm³)
Density Repeatability
- ± 0.02 kg/m³ (± 0.00002 g/cm³)
Flow Rate Indication
- ± 5 % of Reading with 10:1 Turndown
Integral Temperature Sensor
- Class A RTD - with ISO 17025 Accredited, Traceable Calibration
Process Pressure Effect (Corrected)
- ± 0.003 kg/m³ Per 6.9 bar (± 0.021 kg/m³ Per 100 psi)
Process Temperature Effect (Corrected)
- ± 0.005 kg/m³ Per 38° C (± 0.278 kg/m³ Per 100° F)
Documentation
Product Manuals
Brochure
Certifications & Approvals
Document
- Accurate custody transfer and batch measurement with density pdf 303 KB
- ATEX Certificate CDM 100 pdf 2.2 MB
- Cartiere del Garda reduces cost and waste in coatings applic pdf 589 KB
- CDM NORSOK Gudiance Note pdf 695 KB
- CDM100 Measuring Instruments Directive Supplement pdf 1.5 MB
- CDM100 Meters with Foundation Fieldbus Configuration and Use pdf 2.8 MB
- CDM100 MID OIML Certificate of Conformity pdf 577 KB
- Compact Density Meter 100 Retrofit pdf 3.2 MB
- Corrosion Guide for Flow Density and Viscosity Meters pdf 907 KB
- CSA Certificate for CDM100 FDM FVM and HFVM pdf 128 KB
- Density and Concentration Meters pdf 922 KB
- Density Meters to Liquid Hydrocarbon Measurement pdf 805 KB
- EU DECLARATION CDM SENSOR pdf 608 KB
- EU DECLARATION OF CONFORMITY CDM LIQUID DENSITY MERTER pdf 330 KB
- Technical Overview and Specification Summary pdf 4.5 MB