Yodify Product Library
Instantly add the HD2001 Series – T, RH, Pressure and Air Speed Transmitters and 50,000+ more product series to your store or catalog
Book Your Demo and See How
or create your store
HD2001 Series – T, RH, Pressure and Air Speed Transmitters
Brand: Delta OHM S.R.L.The devices of the HD2001 series…, according to the models, measure temperature, relative humidity, barometric pressure and air speed by hot-wire probes.
All the models are provided with RS232C or RS485 serial output and the management of more than one device connected to a network.
In addition, all the models are fitted with an open collector type low-activated configurable alarm output.
The HD2001.1 model has three configurable analog outputs: 4…20 mA or 0…20 mA current output, or 0…10Vdc or 2…10Vdc voltage output.
The choice of output type is made by means of the jumpers set on the board.
Wind speed measurement is detected by the HD2001.2 model with a hot-wire probe set on the upper part of the instrument.
The large display with dual indication on all models allows one of the process variables on the first line and the temperature on the second line, to be displayed.
1 - 5 of 5 Part Numbers
HD2001.1TC1.10
HD2001.1TC1.2
HD2001.1TC1.5
HD2001.1TV1
HD2001.2TV1
1 - 5 of 5 Part Numbers
Specifications
Brand
Temperature Sensor
- NTC 10kΩ
Temperature Working Range
- -20 … +80° C
Temperature Accuracy
- ± 0.3° C in the range 0…+70° C
- ± 0.4° C elsewhere
Humidity %RH Working Range
- 0…100 %RH
Humidity TD Working Range
- -20 … +80° C
Humidity Accuracy
- ± 1.5 %RH (0…90 % RH), ± 2 %RH (elsewhere) For T=15…35° C
- ± (1.5+1.5 % of the measure) %RH in the remaining temperature range
Pressure Working Range
- 600…1100 mbar - 600…1100hPa - 60.0…110.0kPa
Pressure Accuracy
- ± 0.5 mbar @25° C
Pressure Temperature Drift
- < 1 %f.s. (zero), < 1 %f.s. (span), From -20° C to +80° C
Pressure Long-Term Stability
- < 1 mbar/year at 20° C
Pressure Fluid Contacting the Membrane
- Air - Non corrosive gas - No liquids
Wind Speed Type of Sensor
- Hot-wire
Wind Speed Working Range
- 0…5 m/s
Wind Speed °C Working Range
- -20 … +80° C
Wind Speed Accuracy
- ± 0.15 m/s @25° C
Communications Type
- RS-232C and Multidrop RS-485
Communications Maximum Baud Rate
- 9600 baud
Alarm Type of Output
- Open collector (low-activated)
Alarm Maximum Voltage
- 30 Vdc
Alarm Maximum Power
- 200 mW
Variables
- Temperature, %RH, dew point TD, barometric pressure, pressure drop and wind speed
Analog Type of Outputs
- 4...20 mA 0...20 mA 0...10 Vdc 2 … 10 Vdc
Analog Load Resistance
- Current output: 500 Ω max Voltage output: 100kΩ min
Analog Resolution
- 16bit
Power
- 24 Vac ± 10 % 50…60 Hz (230 Vac ± 10 % on request)
Software
- DeltaMet8
Environmental Conditions Temperature Range
- -20 … +80° C
Environmental Conditions Humidity Range
- 0…90 %RH - (without condensation)
Environmental Conditions Protection Degree
- Electronic IP67