Yodify Product Library
Instantly add the 9033X Sodium Analyzer and 50,000+ more product series to your store or catalog  
Book Your Demo and See How
or create your store

Waltron’s 9033X Sodium Analyzer delivers exceptional results in a streamlined low-cost package. 

An enhanced color touch screen user interface makes the 9033X easy to operate and easy to maintain. The 9033X Sodium Analyzer can be used in a variety of ultrapure water applications. 

Common measuring points in a boiler system for online sodium measurement include: make-up water, condensate, boiler feedwater and saturated steam/main steam. 

The Waltron 9033X Sodium Analyzer comes with true fully automatic calibration and offers many benefits in terms of ease-of-use and increased functionality.
Features
  • Large user-friendly color touch screen interface
  • Wide analysis range: 0.1 ppb to 10 ppm
  • Multi-stream available
  • Calibration: automatic 2-point or single point; user may select weekly, bi-monthly, or monthly calibration frequency as well as day and time
  • Hassle-free maintenance: calibration standards and electrode refill can last up to 3 months 
  • Low sample flow indication (optional)
  • Quick connect electrodes and probe positioning clamps
  • Automatic reference electrode electrolyte refill system (standard in US and optional elsewhere)
  • Grab sample capability
  • Two 4-20 mA isolated current outputs
  • Internal data logger set to store readings every 10 minutes for 30 days
  • External data logger can be set for every 1 minute (optional)
  • Automatically logs last 10 alarms and results of last 10 calibrations
  • Reagents: Ammonium Hydroxide (standard) or Di-isopropylamine (optional)

Specifications

Accuracy

  • ± 5 % of Reading or ± 0.1 ppb, whichever is greater within ± 5° C of Calibration Temperature

Ambient Temperature Range

  • 0° to 55° C (32° to 131° F)

Communication

  • Modbus RTU
  • RS-232
  • RS-485

Environmental Protection

  • IP65 (NEMA 4X): Transmitter

Materials of Construction

  • Transmitter: High Strength ABS

Measuring Range

  • 0.10 ppb to 10 ppm

Mounting

  • Horizontal
  • Two Bolt Notches at Bottom
  • Two Key-Style Holes at Top

Output

  • 4-20 mA

Power Supply

  • 90-250 Vac

Pressure Range

  • psig: 5 to 30 (kPa: 34.47 to 206.8, bar: 0.34 to 2.07)

Repeatability

  • Reproducibility: ± 5 % of Reading or ± 0.1 ppb, whichever is greater at Constant Temperature

Response Time

  • 90 % of 10 to 1 ppb Step: 3 minutes
  • 90 % of 100 to 1 ppb Step: 5 minutes

Temperature Range

  • Sample Temperature: 5° to 55° C (41° to 131° F)

Tube Connection

  • Drain: 3/8” (0.953 cm) O.D. Tube
  • Sample Inlet: 1/4” (0.635 cm) O.D. Tube, Swagelok

Wetted Materials

  • PVC
  • Stainless Steel

Instantly populate your store or catalog 
CREATE YOUR SITE
or book a demo with the Yodify Team
show more