Yodify Product Library
Instantly add the Model 111P Percent Oxygen Analyzer and 50,000+ more product series to your store or catalog
Book Your Demo and See How
or create your store
Model 111P Percent Oxygen Analyzer
Brand: Advanced Micro Instruments, IncThe AMI 111 series provide the ideal solution to percent oxygen measurement in non-flammable gases. \\
Most portable percent oxygen analyzers use replaceable electrochemical sensors, which suffer from drift, temperature and barometric pressure sensitivity, and require frequent replacement.
This series of analyzers uses a non-depleting zirconium oxide sensor to measure.
The sensor requires no support gases, and generates an almost perfectly stable reading with minimal drift for months or years.
It is not significantly affected by temperature changes, or by barometric pressure changes.
It will last for many years without being replaced.
When you need an accurate oxygen measurement in an inert background, this series is unique.
Features- Single measurement range; four datalog ranges selectable
- High resolution 3-½ digit LCD
- RFI protected
- Data logger
- USB virtual comport communication for advanced features
- Power requirements: 115/230 Vac to 12 Vdc battery charger
- Low minimum detection limit 0.01% oxygen (25% version)
Specifications
Approvals & Certifications
- CE
- UL (EN 61326-1 2006, EN 61010-1)
Measuring Range
- Optional Range: 0 to 50 %
- Optional Range: 0 to 95 %
- Single Measurement Range: 0 to 25 %
O-Ring Material
- Viton
Power Requirement
- 115/230 Vac to 12 Vdc Battery Charger
Pressure Range
- Inlet Gas Pressure: -14” Hg to 10 psig
Process Connection - Other
- 1/4” Compression Fitting
Repeatability
- ± 0.1 % of Range or ± 0.1 % of Oxygen, whichever is greater
Wetted Materials
- 316 Stainless Steel