Yodify Product Library
Instantly add the 73203 Digital Multimeter and 50,000+ more product series to your store or catalog
Book Your Demo and See How
or create your store
73203 Digital Multimeter
Brand: Yokogawa Electric CorporationDigital multimeters (DMMs) employ an A/D converter with a dual-integration system, which determines the measurement value by converting the input voltage into time using an integration AD converter.
The interval to perform an integral action periodically is referred to as the integral action time.
Features- Compact Size, Ideal for Carrying
- Large Display for Easy Viewing
- Simple Auto Hold Function
- Withstanding Voltage: 3.7 kV AC for 1 Minute
Specifications
Accuracy
- ± (0.3 % of Reading + 1-Digit) - DC Voltage
- ± (0.75 % of Reading + 1-Digit) - Resistance
- ± (0.75 % of Reading + 2-Digits) - Resistance
- ± (0.75 % of Reading + 5-Digits) - AC Voltage
- ± (1 % of Reading + 1-Digit) Testing Current 1 mA or Less - Diode Test
- ± (1 % of Reading + 2-Digits) - DC Current
- ± (2 % of Reading + 1-Digit) - Resistance
- ± (2 % of Reading + 20-Digits), 40 to 500 Hz - AC Current
- ± (2 % of Reading + 2-Digits) - DC Current
- ± (2 % of Reading + 5-Digits), 40 to 500 Hz - AC Current
- ± (2.5 % of Reading + 20-Digits), 40 to 500 Hz - AC Current
- ± (5 % of Reading + 2-Digits) - Resistance
Battery Life
- Approximately 600 Hours (For Continuous DC Voltage Measurement with Alkaline Cells)
Current
- < 0.5 mA - Max.
- < 0.7 µA - Max.
- < 1 mA - Max.
- < 7 µA - Max.
- < 70 µA - Max.
- 10 A, Max.
- 400 mA - Max.
- 400 μA to 10 A - Range (AC Current)
Display
- Digital Display: 4,300 Count Digital Reading
Function
- Auto Hold
- Overvoltage & Current Warning
Humidity
- 70 % Relative Humidity, Non-Condensing or Less at 40° to 50° C (104° to 122° F) - Operating & Storage Humidity
- 80 % Relative Humidity or Less at 0° to 40° C (32° to 104° F) - Operating Humidity
Input
- 500 mA / 250 V (Fuse-Protected) - Input Protection (Capacitor Check)
- 600 V - Max.
- 600 Vrms, Max.
Measuring Range
- 20 NF 200 μF - Capacitor Check
- Digital Display: Approximately 2 times/second
Operating Temperature
- 0° to 50° C (32° to 122° F)
Power Supply
- 2 AAA Dry Cells (LR03 or R03)
Range
- 0.4 to 600 V to AC & DC Voltage
Resistance
- > 10 megaohms, 10 megaohms, 100 megaohms - Input Resistance (DC Voltage)
- > 11 megaohms, 11 megaohms, 11 megaohms, 11 megaohms, < 50 PF - Input Resistance for AC Voltage
- 10 megaohms - Input Resistance (DC Voltage)
- 11 megaohms - Input Resistance (DC Voltage)
- 400 ohms to 40 megaohms - Range
- Buzzer Sounds at 50 ± 20 ohms or Less - Continuity Beeper (Continuity Check)
Response Time
- 0.2 seconds or Less (Buzzer Response) - Continuity Check
- 1 seconds or Less for 400 kiloohms Range - DC Current & Voltage, Resistance, Diode Test & Capacitor Check
- 1.5 seconds or Less for 400 mV Range - DC Voltage
- 15 seconds or Less for 40 megaohms Range - Resistance
- 2 seconds or Less - AC Current & Voltage
- 5 seconds or Less for 4 megaohms Range - Resistance
Storage Temperature
- -20° to 60° C (-4° to 140° F)
Supply Voltage
- < 0.7 V - Open Circuit Voltage (Resistance)
- < 1.0 V - Open Circuit Voltage (Resistance)
- < 3.4 V - Open Circuit Voltage (Resistance)
- 3.7 kVac for 1 Minute - Withstanding Voltage
Voltage Drop
- < 0.04 V/A - AC & DC Current
- < 0.17 mV/μA - AC & DC Current
- < 3 mV/mA - AC & DC Current