Yodify Product Library
Instantly add the HPC40 Series Pressure Calibrator and 50,000+ more product series to your store or catalog  
Book Your Demo and See How
or create your store

HPC40 Series Pressure Calibrator

Brand: Crystal Engineering by Ametek

The HPC40 Series handheld pressure calibrator is the world's first combined pressure and mA loop calibrator to be fully temperature compensated from -20 to 50°C.

You can count on the same accuracy whether measuring pressure, current, voltage, or temperature.
Features
  • 0.035% of Reading Accuracy
  • Gauge, Absolute, and Differential Pressure to 15,000 psi / 1000 bar / 100 MPa 
  • Advanced Simplicity "Non-Menu" Interface 
  • Measure & Source mA with External Loop Power or Internal 24 VDC Power Supply 
  • High Accuracy Thermometer with "True Ohm" Technology 
  • Store & Recall Previously Used Screens
  • Flexible Power Options Including Rechargeable

Specifications

Accuracy (psi - Gauge Pressure)

  • 18 to 28° C -- 0 to 30 % of Range: ± (0.01 % of FS | 30 to 110 % of Range: ± (0.035 % of Reading) | Vacuum*: ± (0.05 % of FS**)
  • -20 to 50° C -- 0 to 30 % of Range: ± (0.015 % of FS | 30 to 110 % of Range: ± (0.050 % of Reading) | Vacuum*: ± (0.05 % of FS**)
  • * Applies to 300 psi and lower ranges only. Vacuum Range = -14.5 psi
  • ** Full Scale is numerical value of the positive pressure range

Accuracy, PSIA (Absolute Pressure with BARO Option)

  • All absolute accuracies are equivalent to the gauge pressure accuracies, except as noted below.
  • 30 psi Range: Gauge Accuracy +0.005 psia
  • 100 psi Range: Gauge Accuracy +0.002 psia

Pressure Sensor (P1 & P2)

  • Wetted Materials: (Wrench Tight) 316 Stainless Steel / (Finger Tight) 316 Stainless Steel and Viton with Internal O-Ring
  • Diaphragm Seal Fluid: Silicone Oil
  • Connection: Crystal CPF Female

Barometric Reference

  • Accuracy: ± 0.00725 psi, ± 0.5 mbar
  • Accuracy specifications include all effects of linearity, hysteresis, repeatability, temperature, and stability within the specified operating temperature range for one year
  • Note: Exposure to environmental extremes of temperature, shock, and/or vibration may warrant a more frequent recertification period

Temperature

  • Accuracy: ± (0.015 % of Reading) + 0.02 Ohm
  • Range: 0 - 400 Ohms
  • Resolution: 0.01 on all scales
  • Units: °C, K, °F, R, Ω
  • Types: 13 Standard and 1 Custom
  • Wiring: 2-, 3-, and 4-Wire Support
  • Connection: Lemo Plug, 1S Series, 304 Insert Configuration

Electrical

  • Connection: 4 mm jacks
  • Maximum Voltage: 45 Vdc

Current (mA) Input

  • Accuracy: ± (0.015 % of Reading + 0.002 mA)
  • mA Range: 0 to 55 mA
  • Percent Range: 0-20, 4-20, 10-50
  • Maximum Allowable Current: 60 mA
  • Resolution: 0.001 mA or 0.01 %
  • Units: mA and %
  • Input Resistance: < 17.2 Ω
  • Voltage Burden @ 20 mA: < 0.35 V
  • Voltage Burden @ 50 mA: < 0.86 V
  • HART Resistor: 250 Ω

Current (mA) Output

  • Accuracy: ± (0.015 of Reading + 0.002 mA)
  • Range: 0 to 25 mA
  • Step Time: 1 to 999 seconds
  • Ramp Time: 5 to 999 seconds

Voltage (VDC) Input

  • Accuracy: ± (0.015 % of Reading + 2 mV)
  • Range: 0 to 30 Vdc
  • Resolution: 0.001 Vdc
  • Input Impedance: > 1 MOhm

Loop Power

  • Fixed Output: 24 Vdc
  • Voltage Output Accuracy: ± 10 %
  • Maximum Output Current: 25 mA

Switch Test

  • Switch Type: Dry Contact
  • Closed State Resistance: 100 KΩ

Display

  • Screen: 320 x 240 pixel Graphical Display
  • Display Rate: 3 readings/second (Standard) / 0 readings/second (Switch Test and Peak hi/lo Modes)

Enclosure

  • Weight: 689 g (24.3 oz.)
  • Rating: IP65
  • Housing: Machined Aluminum, Hard Anodized
  • Keypad and Labels: UV Resistant Silicone

Operating Temperature

  • -20° to 50° C (-4° to 122° F)

Storage Temperature

  • -40° to 75° C (-40° to 167° F)

  • Wind Turbine Testing
    • The HPC40 Series is used to Calibrate Hydraulic Systems found in Wind Turbines