Yodify Product Library
Instantly add the Model ASM Pressure Transducer and 50,000+ more product series to your store or catalog
Book Your Demo and See How
or create your store
Model ASM Pressure Transducer
Brand: Setra Systems, IncSetra’s Model ASM is the highest accuracy transducer for measuring gauge, absolute, compound and vacuum pressure in the AccuSense™ product line.
Its ±0.05% Full Scale accuracy is calibrated using the "End Point Method", which improves linearity when compared to competitive transducers which use the "Best Fit Straight Line Method" of calibration.
The ASM’s calibration is tamper proof by utilizing a SecureCal™ calibration key, which eliminates inadvertent adjustments, while allowing authorized users to adjust the sensor’s calibration coefficients for a true sensor calibration.
The design of the ASM offers class leading overpressure capability and multiple pressure and electrical fittings for a wide range of applications.
Features- End point method linearity
- Low differential pressure ranges
- Low thermal error
- Calibrate using secureCal™ calibration key
- High line pressure capability
- Unidirectional and bidirectional models available
Specifications
Brand
Accuracy
- < ± 0.05 % of Full Scale
- < ± 0.1 % of Full Scale
- < ± 0.10 % of Reading
- Total Error Band: < ± 0.25 % of Full Scale - Thermal
- Total Error Band: < ± 0.5 % of Full Scale - Thermal
Applications
- Engine Test Stands
- Industrial (High Accuracy)
- Manifold Pressure
- Particle Test & Analysis
- Refrigeration Testing
Approvals & Certifications
- CE
- RoHS
Burst Pressure
- psi: 3,000 to 10,000 (kPa: 20,680 to 68,950, bar: 206.8 to 689.5)
Consumption
- < 23 mA
Electrical Connectors
- 2-Wire
- 4-Wire
- 6-Conductor Cable, Pigtail
- 6-Pin Bayonet Connector
Enclosure, Body Material
- Stainless Steel
Environmental Protection
- IP65
- IP66
- NEMA 4X
Excitation Voltage
- 15-30 Vdc (10 Vdc Output)
- 9-30 Vdc (5 Vdc & 4-20 mA Output)
Hysteresis
- < ± 0.03 % of Full Scale Typical
Media
- Liquids & Gases
Non-Linearity
- < ± 0.05 % of Full Scale BFSL - End Point Typical
Non-Repeatability
- < ± 0.02 % of Full Scale Typical
NPT Connection
- 1/4” NPT Female
- 1/4” NPT Male
- 1/8” NPT Female
- 1/8” NPT Male
Operating Temperature
- -40° to 85° C (-40° to 185° F)
Output
- 0-10 Vdc
- 0-5 Vdc
- 4-20 mA
Pressure Cycles
- > 1 Million - Life Cycle Rating
Pressure Range
- bar: -1 to 68.9
- kPa: -101.4 to 6,895
- psi: -14.7 to 1,000
Pressure Type
- Absolute
- Compound Pressure
- Gauge
- Vacuum
Response Time
- Current Output: < 80 milliseconds Pressure Input (From 100 % to 10 % of Pressure Range)
- Voltage Output: < 10 milliseconds Pressure Input (From 100 % to 10 % of Pressure Range)
SAE Connection
- 7/16”-20 SAE Male
Span Effect
- Span Setting Tolerance: < ± 0.01 % of Full Scale
- Span Setting Tolerance: < ± 0.05 % of Full Scale
Stability
- 0.1 % of Full Scale Per Year Typical
Storage Temperature
- -40° to 85° C (-40° to 185° F)
Temperature Range
- Calibrated: -20° to 60° C (-4° to 140° F)
Update Time
- 15 minutes within ± 0.02 % of Full Scale
Vibration
- 10 G from 1,000 to 2,000 Hz
Wetted Materials
- 630 Stainless Steel
Zero Offset
- < 0.05 %/G (Ranges ≥ psi: 100)
- < 0.1 %/G (Ranges ≤ psi: 50)
- Tolerance: < ± 0.01 % of Full Scale
- Tolerance: < ± 0.05 % of Full Scale