Yodify Product Library
Instantly add the RVP4500 Series Reid Vapor Pressure Analyzer and 50,000+ more product series to your store or catalog  
Book Your Demo and See How
or create your store

RVP4500 Series Reid Vapor Pressure Analyzer

Brand: ABB Group

Increase your gasoline blending profits with process Reid Vapor Pressure measurements and automatic air saturation of the sample.

The ABB RVP4500 Series accurately measures Reid Vapor Pressure throughout the gasoline range and its feedstocks. 

The RVP4500 Series offers a simple, reliable, compact, single cell design that requires minimal operator interface, minimal maintenance and no calibration. 

It validates every analysis with self-diagnostics.
Features
  • Microprocessor based on single board computer
  • Single cell design
  • Approximate cycle time 7.5 minutes
  • Measurement range 0 to 20 psi
  • Low sample volume  2.5 ml
  • Explosion proof
  • Repeatable, reproducible, and accurate
  • Require minimal operator interface
  • Instrument checks operation via self-diagnostics
  • Small compact design

1 - 6 of 6 Part Numbers
RVP4500
RVP4500
  • Product - 0-20 psi
  • Measuring Gas - Gasoline
RVP4501
RVP4501
  • Product - 0-90 psi
  • Measuring Gas - Gasoline
RVP4503
RVP4503
  • Product - 0-30 psi
RVP4510
RVP4510
  • Product - 0-25 psi
RVP4540
RVP4540
  • Product - 0-225 psi
RVP4550
RVP4550
  • Product - 0-20 psi
1 - 6 of 6 Part Numbers

Specifications

Brand

Environmental Protection

  • IP66 (NEMA 4 Equivalent)

Ambient Temperature

  • 0° to 32° C (32° to 90° F)
  • 0° to 40° C (32° to 104° F) with Vortex option

Humidity

  • 95 % Relative Humidity, Non-Condensing

Mounting

  • Wall Mount
  • Floor - Optional dolly

EMI / RFI

  • Conform to Class A industrial environment

Certifications & Approvals

  • CSA/NRTL
  • ATEX
  • IECEx

Power Supply

  • 100-240 Vac, 50 to 60 Hz

Power Consumption

  • 150 W

Process Connection

  • 1/4” (6.4 mm) tube, minimum

Supply Pressure

  • 414 kPa (60 psig) Min.

Measurement Range

  • 0 to 20 psi

Cycle Time

  • Approximately 7.5 minutes

  • ASTM Method D5482 (off-line mode)
    • The ASTM Method D5482 was developed for this analyzer
    • Allows results from process measurements to be accepted by the regulatory agencies
  • ASTM Method D323
    • The measurement results correlate with the ASTM D323 with no correction factors or calibration required.
  • Other Applications
    • The measurement results correlate with the ASTM D323 with no correction factors or calibration required.
    • Gasoline Blending
    • Measuring lights in de-butanizer tower bottoms
    • Shale Gas