Yodify Product Library
Instantly add the DM-702-NH₃ Ammonia Sensor and 50,000+ more product series to your store or catalog  
Book Your Demo and See How
or create your store

DM-702-NH₃ Ammonia Sensor

Brand: Detcon Inc

Detcon Model DM-702-NH₃ is a non-intrusive “Smart” sensor designed to detect and monitor Ammonia in air over the range of 0-100 ppm using electrochemical sensor technology. 

The intelligent plug-in, field replaceable cell provides automatic recognition of gas type and range, and features over-sized gold-plated connections that help prevent corrosion. 

The Model 700’s rugged framework includes an intrinsically safe electro-polished 316 stainless steel housing with fully encapsulated electronics and dual layer surge protection. 

This innovative design virtually eliminates sensor failure due to water ingress, corrosion, vibration, and transient spikes. 

A primary feature of the Model 700 is embedded intuitive software that simplifies operator interface by guiding the user through routine calibration, configuration, and fault diagnostic functions using a built-in alpha/numeric display. 

The Model 700 is equipped with standard Analog 4-20 mA, and Modbus™ RS-485 outputs. 

Among its unique features is a wireless option that can be used with Detcon’s SmartWireless® product line. 

Additional integration options include a Remote Alarm Module (RAM), HART, and Foundation Fieldbus.

1 - 3 of 3 Part Numbers
967-505090-100
967-505090-100
  • Junction Box - No Junction Box
967-505091-100
967-505091-100
  • Junction Box - Aluminum
967-505094-100
967-505094-100
  • Junction Box - 316 Stainless Steel
1 - 3 of 3 Part Numbers

Specifications

Brand

Accuracy

  • ± 2 % of Full Scale

Approvals & Certifications

  • ATEX II 2 GD Ib IIC T4 Gb (Tamb: -40° to 50° C)
  • cCSAus
  • CE
  • CSA: Class I, Div. 1, Groups B, C, D (Tamb: -40° to 50° C), Class I, Zone 1, Group IIC
  • NRTL: Class I, Div. 1, Groups B, C, D (Tamb: -40° to 50° C), Class I, Zone 1, Group IIC
  • SIL 2 Certified to IEC 61508

Cable Length

  • Maximum

Circuit Protection

  • EMI/RFI Immunity - I/O Protection
  • Miswiring - I/O Protection
  • Over-Voltage - I/O Protection

Communication

  • FOUNDATION Fieldbus
  • HART
  • Modbus RTU
  • RS-485

Consumption

  • < 0.75 W at 30 mA - Normal Operation for Power Consumption
  • 1.2 W at 50 mA - Max.

Current

  • 500 mA at 24 V Typical - Inrush Current

Display

  • LED Display (with Antiglare Cover)

Electrical Connectors

  • 14 AWG (2.5 mm²)
  • 2-Wire Twisted-Pair Shielded Cable
  • 3-Wire Shielded Cable

Enclosure, Body Material

  • 316 Stainless Steel

Environmental Protection

  • Corrosion-Proof
  • Explosion-proof
  • IP66
  • NEMA 4X
  • Vibration-Proof
  • Water-Proof

Humidity

  • 15 to 90 % Relative Humidity, Non-Condensing

Indication

  • Status Indicators: 4-Digit LED Display with Gas Concentration Full-script Menu Prompts for AutoSpan, Set-Up Options, & Fault Reporting

Input Power

  • 11-30 Vdc

Loop Resistance

  • 1,000 ohms, Max.

Measuring Range

  • 0 to 100 ppm

Operating Pressure

  • Atmospheric ± 10 %

Operating Temperature

  • -40° to 50° C (-40° to 122° F)

Output

  • 0 mA - All Fault Diagnostics
  • 2 mA - In-Calibration
  • 22 mA - Over-Range Condition
  • 4-20 mA - 0 to 100 % of Full Scale
  • 4-20 mA DC - Linear

Repeatability

  • ± 2 % of Full Scale

Response Time

  • T90 < 90 seconds

RFI/EMI Protection

  • Complies with EN 61000

Sensor, Probe Type

  • 3-Electrode Electrochemical Cell
  • Continuous Diffusion / Absorption
  • Plug-In Field Replaceable Intelligent Typical

Span Effect

  • < 2 % Signal Loss Per Month

Storage Temperature

  • -35° to 55° C (-31° to 131° F)

  • Chemical Plants
  • Food & Beverage
  • Oil & Gas Applications
  • Pulp & Paper
  • Refineries
  • Steel Mills
  • Utilities
  • Wastewater Treatment Plants

Instantly populate your store or catalog 
CREATE YOUR SITE
or book a demo with the Yodify Team
show more