Yodify Product Library
Instantly add the 5100/5105-ISO Tension & Compression Load Cell and 50,000+ more product series to your store or catalog
Book Your Demo and See How
or create your store
5100/5105-ISO Tension & Compression Load Cell
Brand: Sensy - Sensors & Synergy S.A.Standard reference force transducer specially designed according to the norm ISO 376(1) (Class "1", "05" and "00").
Features- Broad range of capacities (Up to 30 MN)
- Complete range of load accessories
- Protection: IP 65
- Material: nickel plated alloy steel
- Options:
- Digital output RS-232C or RS-485
- Custom made dimensions
- ASTM E74 compliant
Specifications
Accuracy
- Class 00
- Class 05
- Class 1
Applications
- Control of Testing Machines According to ISO 7500
Approvals & Certifications
- ATEX
- CE
Cable Length
- 10 to 39 feet (3 to 12 metres)
Capacity
- 1 to 3,000 T
Communication
- RS-232
- RS-485
Electrical Connectors
- Female-Male Connector DIN 45322
Enclosure, Body Material
- Nickel-Plated Alloy Steel - Model 5105
- Stainless Steel - Model 5100
Environmental Protection
- IP65
- IP66
Excitation Voltage
- 10 Vdc - Reference Excitation Voltage
- 3-12 Vdc - Nominal Range
Insulation Resistance
- > 5,000 megaohms at 50 V
Load
- > 300 % of Full Scale - Breaking Load
Load Limit
- 110 % of Full Scale - Safe Load Limit
Metric Connection
- M100 x 4 Male
- M125 x 4 Male
- M160 x 6 Male
- M200 x 6 Male
- M24 x 2 Male
- M250 x 6 Male
- M30 x 2 Male
- M330 x 6 Male
- M360 x 6 Male
- M45 x 3 Male
- M64 x 4 Male from 0.75 to 1 MN
- M90 x 4 Male
Repeatability
- Relative Error with Rotation: < ± 0.05 % R.O to < ± 0.20 % R.O
- Relative Error without Rotation: < ± 0.025 % R.O to < ± 0.10 % R.O
Resistance
- 350 to 702 ± 2 ohms - Input & Output Resistance
Sensitivity
- 1.5 mV/V - Nominal Sensitivity
Service Temperature
- -30° to 70° C (-22° to 158° F)
Storage Temperature
- -50° to 85° C (-58° to 185° F)
Temperature Coefficient
- Sensitivity: Class 00 < ± 0.015 % of Full Scale/10° C (18° F)
- Sensitivity: Class 1, & Class 05 < ± 0.035 % of Full Scale/10° C (18° F)
- Zero Signal: Class 00 < ± 0.023 % of Full Scale/10° C (18° F)
- Zero Signal: Class 1, & Class 05 < ± 0.03 % of Full Scale/10° C (18° F)
Temperature Range
- -10° to 45° C (14° to 113° F) - Nominal Temperature Range
- 20° C (68° F) - Reference Temperature
Zero Balance
- ± 0.02 mV/V
Zero Offset
- Zero Shift after Loading: < ± 0.05 % of Full Scale (Class 1) to < ± 0.012 % of Full Scale (Class 00)