The EX100/200-Series are adjustable depth insertion magmeters that fit 3” to 48” pipe (up to 72” optional). The complete lack of moving parts of the EX100/200-Series is the source of its reliability. Brass and stainless steel models withstand a variety of temperature, pressure, and chemical conditions. The EX-Series has no rotor to stop turning in dirty water and there are no bearings to wear out. Like all magmeters, when used in chemical injection applications, these meters should be installed upstream of the chemical line (or far enough downstream to allow complete mixing of fluids before the meter). 

Adapters mate with standard 1-1/2” (11x/21x) or 2” (15x/25x) FNPT threaded fittings such as saddles and weldolets which may be purchased either locally or from Seametrics. A rapidly reversing magnetic field is produced in the lower housing. As the fluid moves through this field, a voltage is generated that is measured and translated into a frequency signal proportional to flow rate. This square wave signal can be sent directly to a PLC or other control or can be converted using any of the Seametrics family of indicators and converters. A modular system of electronics can be installed directly on the flow sensor or mounted remotely. 

The FT430 (externally powered with pulse) and the FT440 (loop powered), both provide digital rate and total displays, as well as a programmable pulse; the FT440 also provides a 4-20 mA analog output. The “hot-tap” models (EX15x/25x) can be installed or serviced without shutting down the line by means of a 2” full-port isolation valve that comes with a nipple for installation on the pipe fitting; a bronze ball valve is standard on both models, with a 316 stainless steel valve optional if needed. In most circumstances, no special tool is required.
Features
  • Isolation valve allows installation and removal under pressure
  • Useful in large water utility pipe and other applications where shutdown is not an option
  • EX150 fits pipe from 3″ to 10″
  • EX250 fits pipe from 10″ to 48″
  • Meter extends only about 1/8 of pipe diameter, minimizing potential for clogging with debris
  • In most applications, installation tool not necessary due to small tube diameter
  • Available in brass or 316 stainless steel
  • Standard valve material is brass, with stainless optional
  • Certified to NSF/ANSI 61 Standard (stainless version)

Specifications

Pipe Size

  • 3” to 48” (up to 72” optional)

Power

  • Full Power: 12-24 Vdc, 250 mA
  • Low Power: 12-24 Vdc, 40 mA average with 250 mA peaks

Materials Housing

  • Powder-coated cast aluminum

Tubing/Fitting/Sensor Housing

  • Brass or 316 Stainless Steel

Electrodes

  • Hastelloy

Electrode Cap

  • PVDF

O-Ring (15x/25x only)

  • EPDM

Valve Assembly (15x/25x only

  • Bronze ball valve, both models (Stainless steel valve optional for stainless steel model)

Fitting Size Required

  • 11x/21x: 1.5” FNPT
  • 15x/25x: 2” FNPT

Maximum Pressure

  • 200 psi (14 bar)

Temperature Ambient

  • 0° to 160° F (-17° to 72° C)

Fluid

  • 32° to 200° F (0° to 93° C)

Minimum Conductivity

  • 20 microSiemens/cm

Flow Velocity

  • 0.28 - 20 ft/sec (0.08 - 6.09 m/sec)

Accuracy

  • ± 1 % of full scale

Output

  • Square wave pulse, opto-Isolated, 500 Hz @ 20 ft/sec 6 mA max., 30 Vdc forward flow standard; reverse flow optional

Empty Pipe Detection

  • Software, defaults to zero flow

Cable

  • Standard 18’ (6 m), #22 shielded twisted pair, 4-conn. Max. cable run at 24 Vdc = 1000’ (300 m); at 12 Vdc = 500’ (150 m). For other circumstances, contact the factory.

Regulatory

  • Mark (applies to full power sensor only configuration and full power FT 430/440 mounted configurations). Certified to NSF/ANSI standard 61 and NSF 372 (Stainless only)