Brand: MTL InstrumentsThe K6050AP analyser has been designed specifically for use in power stations.
Modern turbo-generators produce a great amount of heat and the most efficient way of removing this heat is by enveloping the parts in hydrogen, because hydrogen has seven times the cooling capacity of air.
It is possible for air to leak into the system, so it is essential to monitor the hydrogen purity for both efficiency and safety - as a hydrogen/air mixture will produce a potentially explosive atmosphere.
When maintenance needs to be carried out, the atmosphere inside the turbo-generator has to be replaced with air.
This cannot be a direct replacement because of the explosive mixture that would be produced in the process.
Instead, an inert gas, usually carbon dioxide, is used to purge the hydrogen, then air is introduced to purge the carbon dioxide.
The same procedure must be followed in reverse when replacing the hydrogen atmosphere.
Features- Proven Sensor Technology
- Measures All Gas Purge Stages
- Weatherproof to IP65
- Reliable, Simple to Use
- Data Logger Output
- Ideal Backup for Fixed Analyzers
Specifications
Brand
Display
- Dot-matrix LCD showing two or four lines of alphanumeric characters
Ranges
- 80 to 100 % H2 in Air
- 0 to 100 % H₂ in CO
- 0 to 100 % CO₂ in Air
Typical Accuracy
- ± 1 % of span
Stability
- Better than 1 % FSD/month
Resolution
- 0.1 % Hydrogen ranges
- 0.5 % CO₂ in Air
Typical Speed of Response
- T90 < 5 seconds
Sample Connections
- Compression fitting suitable for 0.25 inch (6 mm) O.D. tubing
Sample Pressure
- Pump Off: +6 barg Max.
- Pump On: -100 mbarg Mininum
Sample Flow
- 100 to 300 mL/min for optimum performance
Sample Temperature
- -5° to +40° C (Non-Condensing)
Outputs
- 0 to 1 V analogue
Ambient Temperature
- -5° to +40° C
Battery Capacity
- 6 hours
Supplied Battery Chargers
- Mains: 110/240 Vac
- Vehicle cigarette lighter: 12 Vdc @ 1.25A Max.
- Hydrogen Cooled Turbo-generators
- Hydrogen Cooled Synchronous Compensators
- Analyzers Available for CO2 or N2 Purging