Yodify Product Library
Instantly add the OPTITEMP TT 60 C/R Universal Temperature Transmitter and 50,000+ more product series to your store or catalog
Book Your Demo and See How
or create your store
OPTITEMP TT 60 C/R Universal Temperature Transmitter
Brand: KrohneThe OPTITEMP TT 60 C/R series feature extremely accurate measurements as well as sensor and process information.
With five connection terminals, new functions such as two redundant Pt100 in 3-wire connections are included.
Programming is done either using the Krohne / Inor software ProfiSoft via PC or using the Profibus network.
The OPTITEMP TT 60 series consists of 2 different versions in terms of mounting. The TT 60 R is the the rail-mount version whereas the TT 60 C is primarily intended to be mounted in a "B connection head" or larger in accordance with DIN EN 50446.
Alternatively, you can also mount the in-head version on a DIN rail according to EN 60715 with the help of an optionally available rail mounting kit.
Features- Up to 125 transmitters in one Profibus network
- Profile version 3.0, A and B
- Completely universal, linearized and separate
- Input: resistance thermometers, thermocouples, mV and Ohms
- Double Pt100 in 3-wire connection and TC input
- Several outputs: input value from channel 1 and channel 2, process value to scale, redundancy with double sensor elements, arithmetic functions (difference, mean value, minimum and maximum)
- Easy wiring, large center hole
- Sensor error correction
- 50-point linearization: each sensor can be adapted
- Excellent sensor monitoring functions such as: sensor break, sensor short circuit, detection of low sensor isolation, sensor drift
- Robust design, tested for 5g vibration
- Integrated in the siemens PDM system
Specifications
Brand
Accuracy
- 0.1° C (0.18° F) Typical - Pt100
Ambient Temperature Range
- -20° to 70° C (-4° to 158° F) - TT 60 R
- -40° to 85° C (-40° to 185° F) - TT 60 C
Applications
- Chemical Industry
- Food & Beverage
- Iron, Steel & Metal Industries
- Oil & Gas Applications
- Pharmaceutical
- Power Plants
- Pulp & Paper
Approvals & Certifications
- CE Ex II 3 G EEx NL IIC, 0102 Ex II 1 G Ex ia IIC
- Demko 06-ATEX-141337X
Capacitance
- 1 NF - Bus Connection (Input)
- 10 NF - Communications Connection (Input)
- 1000 NF - Sensor Connection (Output for TT 60 C Version)
- 166 NF - Sensor Connection (Output for TT 60 C Ex-Version)
Communication
- PROFIBUS PA
Consumption
- 11 mA - Power Consumption
Electrical Connectors
- Shielded Cable
Environmental Protection
- IP20
- IP54
Inductance
- 1 µH - Communications Connection (Input)
- 10 µH - Bus Connection (Input)
- 750 MH - Sensor Connection (Output)
Input
- -10 to 1,000 mV
- 17.5 Vdc, Max.
- 380 mA - Max.
- 450 mA - Max.
- 8 Vdc, Max.
Input Power
- 5.32 W, Max.
- 800 mW, Max.
Isolation
- 1,500 Vac, 1 Minute - Galvanic
Mounting
- DIN Form B Connection Head - According to DIN 43729
- M4 Screw
- Rail Mounting - According to EN 50020, 35 mm
Output
- 17.5 Vdc - Sensor Connection
- 56 mA - Sensor Connection
Output Power
- 244 mW - Sensor Connection
Power Supply
- 9-17.5 Vdc - TT 60 C
- 9-32 Vdc - TT 60 R
Resistance
- 0 to 4,000 ohms - Input Resistance
RTD Type
- Ni100
- Ni1000
- Ni120
- Ni50
- Potentiometer
- Pt10
- Pt100
- Pt1000
- Pt200
- Pt50
- Pt500
Sensor, Probe Connection
- 2-Wire
- 3-Wire
- 4-Wire
Sensor, Probe Type
- B Thermocouple
- C Thermocouple
- D Thermocouple
- E Thermocouple
- G Thermocouple
- J Thermocouple
- K Thermocouple
- L Thermocouple
- N Thermocouple
- RTD
- R Thermocouple
- S Thermocouple
- T Thermocouple
- U Thermocouple
Temperature Range
- -200° to 250° C (-328° to 482° F) - Ni50, Ni100, Ni120, Ni1000
- -200° to 850° C (-328° to 1,562° F) - Pt10, Pt50, Pt100, Pt200, Pt500, Pt1000