Fluke Calibration 9173 Field Metrology Well
Brand: Fluke CorporationEvery once in a while, a new product comes around that changes the rules.
It happened when we introduced handheld dry-wells.
It happened when we introduced Micro-Baths.
Now we’ve combined bath-level performance with dry-well functionality and legitimate reference thermometry to create Metrology Wells.
With groundbreaking new proprietary electronics from Fluke Calibration's (patents pending), Metrology Wells let you bring lab-quality performance into whatever field environment you might work in.
New analog and digital control techniques provide stability as good as ±0.005 °C. And with dual-zone control, axial (or “vertical”) uniformity is as good as ±0.02 °C over a 60 mm (2.36 in) zone. (That’s 60 mm!)
Such performance doesn’t exist anywhere else outside of fluid baths.
In short, there are six critical components of performance in an industrial heat source: calibrated display accuracy, stability, axial (vertical) uniformity, radial (well-to-well) uniformity, impact from loading, and hysteresis.
We added a seventh in the form of a legitimate reference thermometer input and created an entirely new product category: Metrology Wells.
Features- Best-performing industrial heat sources (accuracy, stability, uniformity) in the world
- Immersion depth to 203 mm (8 in)
- Optional ITS-90 reference input reads PRTs to ±0.006 °C
- Temperature range from –45 °C to 700 °C
1 - 12 of 12 Part Numbers
Fluke 9173-A
Fluke 9173-A-R
Fluke 9173-B
Fluke 9173-B-R
Fluke 9173-C
Fluke 9173-C-R
Fluke 9173-D
Fluke 9173-D-R
Fluke 9173-E
Fluke 9173-E-R
Fluke 9173-F
Fluke 9173-F-R
1 - 12 of 12 Part Numbers
Specifications
Brand
Range at 23° C Ambient
- 50° to 700° C
Display Accuracy
- ± 0.2° C: 50° to 425° C
- ± 0.25° C: 425° to 660° C
Stability
- ± 0.005° C: 50° to 100° C
- ± 0.01° C: 100° to 425° C
- ± 0.03° C: 425° to 700° C
Axial Uniformity
- ± 0.1° C: 50° to 100° C
- ± 0.25° C: 100° to 425° C
- ± 0.4° C: 425° to 700° C
Radial Uniformity
- ± 0.01° C: 50° to 100° C
- ± 0.025° C: 100° to 425° C
- ± 0.04° C: 425° to 700° C
Loading Effect
- ± 0.02° C at 425° C
- ± 0.04° C at 700° C
Well Depth
- 203 mm
Resolution
- 0.001° C
Computer Interface
- RS-232 Interface
Temperature Range
- -200° to 962° C
- -328° to 1764° F
Resistance Range
- 0 Ω to 400 Ω, auto-ranging
Resistance Accuracy
- 0 Ω to 20 Ω: 0.0005 W
- 20 Ω to 400 Ω: 25 ppm
Resistance Resolution
- 0 Ω to 20 Ω: 0.0001 Ω
- 20 Ω to 400 Ω: 0.001 Ω