Yodify Product Library
Instantly add the 2792A Series Standard Resistor and 50,000+ more product series to your store or catalog
Book Your Demo and See How
or create your store
2792A Series Standard Resistor
Brand: Yokogawa Electric CorporationThe 2972A Series of standard resistors are metal foil resistors, while the previous 2972 Series were winding resistors.
The error range of the 2972A Series is much smaller than those of the previous 2972 Series, as demonstrated by the resistance temperature coefficient curves.
Precision temperature control equipment such as oil baths, are not needed for measurement and calibration. The 2972A can do these in an air environment.
Features- Traced to the National Standard for high accuracy; test (calibrated) accuracy of ±5 ppm
- Resistance temperature coefficient. Excellent temperature characteristics in the range of 0° to 50° C (32° to 122° F) resistance temperature coefficient less than ±2.5 ppm / ° C
- A variety of models. Eight models with nominal resistance values ranging between 0.001 Ohm and 10 KiloOhms
- Precision temperature control equipment, such as an oil bath, not needed for calibration due to marked improvement in resistance temperature coefficient
1 - 8 of 8 Part Numbers
2792A01
- Nominal Value - 0.001 Ohm
2792A02
- Nominal Value - 0.01 Ohm
2792A03
- Nominal Value - 0.1 Ohm
2792A04
- Nominal Value - 1 Ohm
2792A05
- Nominal Value - 10 Ohms
2792A06
- Nominal Value - 100 Ohms
2792A07
- Nominal Value - 1 KiloOhm
2792A08
- Nominal Value - 10 KiloOhms
1 - 8 of 8 Part Numbers
Specifications
Accuracy
- ± 30 to ± 100 ppm at 23° C (73.4° F) ± 2° C (± 3.6° F)
- Test (Calibrated) Accuracy: ± 5 ppm
Current
- Maximum
Humidity
- 20 to 80 % Relative Humidity
Insulation Resistance
- More Than 1,000 megaohms at 500 Vdc
Operating Temperature
- 0° to 50° C (32° to 122° F)
Power
- Maximum
- Power Characteristics: ± 100 ppm/W
Resistance
- Nominal Resistance: 0.001 ohm to 10 kiloohms
Storage Temperature
- -20° to 60° C (-4° to 140° F)
Supply Voltage
- Withstand Voltage: 1.5 kV for One Minute between Measurement Terminal & Casing
Temperature Coefficient
- -5° to ± 15 ppm/°C
- Less than -0.1 ppm/°C²
Thermal Drift
- Drift: ± 30 to ± 100 ppm Per Year