Brand: Yokogawa Electric CorporationThe AQ6375B and AQ6376 are the bench-top optical spectrum analyzers covering the long wavelengths over 2 μm.
They are designed for researchers and engineers who have been struggling with inadequate test equipment to measure in these long wavelength ranges.
The AQ6375B and AQ6376 achieve high speed measurements with high accuracy, resolution and sensitivity, even while providing full analysis features.
Cumbersome calibration steps and the development of external analysis software is no longer required.
Features- Gas purging feature
- Built-in cut filter for high order diffracted light
- Light source for wavelength calibration
- Free space optical input
Specifications
Accuracy
- ± 0.05 Nm (1,520 to 1,580 Nm) - Wavelength Accuracy (AQ6375B)
- ± 0.1 Nm (1,580 to 1,620 Nm) - Wavelength Accuracy (AQ6375B)
- ± 0.5 Nm (Full Range) - Wavelength Accuracy
- ± 1 dB (1,550 Nm, Input Level: -20 dBm, Sensitivity: MID, HIGH1-3) - Level Accuracy
Approvals & Certifications
- EN 61010-1
- GB
Approvals & Certifications
- AS
- BS
- CSA
- EMC Emission: EN 61326-1 Class A, EN 55011 Class A Group 1. Immunity: EN 61326-1 Table 2
- IEC 60825-1 Class I, (Laser)
- IEEE 488.1/488.2
- NBR
- RoHS EN 50581
- UL
- VDE
Communication
- Ethernet
- Ethernet - Electrical Interface
- Ethernet TCP/IP - Remote Control
- GPIB - Electrical Interface, Remote Control
- RS-232
- USB
- USB - Electrical Interface
Connectors
- Fiber Optic Connector - Single Mode (9.5/125 μm), Graded Index (50/125 μm, 62.5/125 μm)
- Optical Connector (FC or SC)
Diameter
- 1/4” (6.35 mm), Nylon Tube - Purge Gas Input/Output Terminals
Display
- 10.4-inch Color LCD (Resolution: 800 x 600)
Electrical Connectors
- Universal Adapter
Function
- Data Logging
- Smoothing
Humidity
- 20 to 80 % Relative Humidity, Non-Condensing
Input Power
- 13 dBm (Per Channel, Full Wavelength Range) - Max.
- 20 dBm - Max.
- 25 dBm (Total Input Power) - Max.
Linearity
- ± 0.05 dB (Input Level: -30 to 10 dBm, Sensitivity: HIGH1-3) - Level Linearity (AQ6375B)
- ± 0.2 dB (Input Level: -30 to 10 dBm, Sensitivity: HIGH1-3) - Level Linearity (AQ6376)
Materials of Construction
- Nylon - Tube
Measuring Range
- -65 to 13 dBm (AQ6376) - Wide Measurable Level Range
- -70 to 20 dBm (AQ6375B) - Wide Measurable Level Range
Operating Temperature
- 5° to 35° C (41° to 95° F)
Polarization
- ± 0.1 dB (1,550 Nm) - Polarization Dependence (AQ6375B)
Port
- Analog Output Port - Electrical Interface
- SVGA Output
- Trigger Input Port
- Trigger Output Port
Power Requirement
- 100-240 Vac, 50/60 Hz
- Approximately 100 VA
Range
- 40 dB (Peak ± 1 Nm, Resolution: 0.1 Nm) - Dynamic Range (AQ6376)
- 45 dB (Peak ± 0.4 Nm, Resolution: 0.05 Nm) - Dynamic Range (AQ6375B)
- 55 dB - Close-in Dynamic Range
- 55 dB (Peak ± 0.8 Nm, Resolution: 0.05 Nm) (1,523 Nm, Sensitivity: HIGH1-3) for (AQ6375B), (Peak ± 2 Nm, Resolution: 0.1 Nm) (1523 Nm, Sensitivity: HIGH1-3) for (AQ6376) - Dynamic Range
Remote Control & Monitoring Software
- Ethernet
Repeatability
- ± 0.015 Nm (1 Minute) - Wavelength Repatability
Resolution
- 0.002 Nm - Min. Sampling Resolution (AQ6375B)
- 0.003 Nm - Min. Sampling Resolution (AQ6376)
- 0.1 to 2 Nm (AQ6376)
- 50 Pm to 2 Nm (AQ6375B)
Sensitivity
- -50 dBm (3,200 to 3,400 Nm) (Sensitivity: HIGH3) - Level Sensitivity (AQ6376)
- -55 dBm (2,200 to 3,200 Nm) - Level Sensitivity (AQ6376)
- -62 dBm (1,300 to 1,500 Nm) (Sensitivity: HIGH3) - Level Sensitivity (AQ6375B)
- -65 dBm (1,500 to 2,200 Nm) - Level Sensitivity (AQ6376)
- -67 dBm (1,500 to 1,800 Nm, 2,200 to 2,400 Nm) - Level Sensitivity (AQ6375B)
- -70 dBm (1,800 to 2,200 Nm) - Level Sensitivity (AQ6375B)
Storage Temperature
- -10° to 50° C (14° to 122° F)
Temperature Range
- 18° to 28° C (64.4° to 82.4° F) - Performance Guarantee Temperature
Update Time
- Minimum 1-Hour (After Warming Up, Optical Alignment Adjustment with Built-in Light Source Required) - Warm-Up Time
Wavelength
- 0.5 to 1,200 Nm (Full Span), 0 Nm - Span (AQ6375B)
- 0.5 to 1,900 Nm (Full Span), 0 Nm - Span (AQ6376)
- 1,200 to 2,400 Nm - AQ6375B
- 1,500 to 3,400 Nm - AQ6376
- Gas Detection & Concentration