Metis M311 & M322 2-Color Pyrometer Series

Brand: Process Sensors Corporation

Highly advanced, full featured 2-color pyrometer series.
Features
  • Fully digital and very fast with response time <1 ms
  • Adjustable or motorized focus optics
  • Small spot sizes from 0.9 mm
  • Laser, color video or thru-lens sighting
  • Dirty window programmable alarm
  • 10-digit matrix display for temperature and IR sensor parameters
  • Push button device configuration or via software
  • 3 versatile configurable inputs or outputs
  • Analog input for external emissivity setting
  • Highest accuracy and repeatability even at high ambient temperatures up to 80° C (176° F) without cooling
  • Temperature ranges between 300° and 3,300° C (572° and 5,972° F)
  • 2 high resolution 16-bit Analog 0-4 to 0-20 mA outputs
  • Serial interfaces RS232 and RS485 (switchable)
  • Optional fieldbus connection: Profinet or Profibus

Specifications

Ambient Temperature Range

  • Fiber Optic Version, Focusable Lens Assembly: -20° to 250° C (-4° to 482° F)
  • Operation: 0° to 80° C (32° to 176° F)

Applications

  • Brazing
  • Ceramic Industry
  • Composites
  • Induction Heating
  • Kilns
  • Metal
  • Metal Pour Streams
  • Nuclear
  • Pipe Bending
  • Research & Development
  • Semiconductor Industry
  • Sintering
  • Steel
  • Vacuum Furnaces
  • Welding

Approvals & Certifications

  • CE EU Directives for Electromagnetic Immunity

Communication

  • FOUNDATION Fieldbus
  • PROFIBUS
  • PROFINET
  • RS-232, Max.
  • RS-485, Max.

Display

  • 10-Digit LED Display (5 mm High) for Temperature or Settings of IR Sensor Parameters

Electrical Contacts

  • Normally Closed
  • Normally Open

Electrical Protection

  • Reverse Polarity

Enclosure, Body Material

  • Aluminum

Environmental Protection

  • IP65

Load

  • Maximum

Output

  • Analog Output Signal: 0-20 mA
  • Analog Output Signal: 4-20 mA

Power Requirement

  • 24 Vdc (18-30 Vdc) - Max.

Range

  • 0.75 to 1.8 µm

Repeatability

  • 0.1 % of Measured Value in °C + 1 K

Resolution

  • Analog Output Signal: 0.0015 % of the Adjusted Temperature (16-bits)
  • Serial Interface & Display: 0.1° C
  • Serial Interface & Display: 0.1° F

Response Time

  • < 1 millisecond (With Dynamical Adaptation at Low Signal Levels), Adjustable 0 to 10 seconds

Span - Minimum

  • 50° C (122° F)

Storage Temperature

  • -20° to 85° C (-4° to 185° F) (The Laser Targeting Light is Deactivated at A Device Temperature from 60° C, the Camera Module from 55° to Prevent Its OverHeating)

Temperature Coefficient

  • From 0° to 10° C & 60° to 80° C: 0.06 %/K
  • From 10° to 60° C: 0.04 %/K

Temperature Range

  • 300° to 3,300° C (572° to 5,972° F)

Uncertainty

  • Full Scale Temperature Above 2500° C: 0.5 % of Measured Value in °C
  • Full Scale Temperature up to 2500° C: 0.3 % of Measured Value in °C + 2 K