PGC5007 Total Sulfur Analyzer
Brand: ABB GroupThe PGC5007 performs a Total sulfur analysis for a range of streams, liquid and vapor, from natural gas to gasoline and diesel.
A fixed volume of sample from the process stream is injected via a sample inject valve.
Air transports the sample into the furnace, where it oxidizes the sample to carbon dioxide, water and sulfur dioxide.
These components are separated using packed columns and pass into the Flame Photometric Detector (FPD), where the trace levels of Total Sulfur are measured.
Specifications
Brand
Environmental Protection
- IP54 (NEMA 3 Equivalent)
Ambient Temperature
- 0° to 50° C (32° to 122° F)
Humidity
- 95 % Relative Humidity, Non-Condensing
Mounting
- Wall: 32 mm (1.3 in.) from wall with brackets
EMI / RFI
- Conforms to Class A industrial environment
Certifications & Approvals
- CN / KO / RU: Ex px de IIB+H₂ T3 - T2, with X-purge Power interlock
- ATEX / IEC: Zone 1: CE 0344, II2G, Ex py de IIB+H₂ T3 - T2, Zone 2: CE, II3G Ex nA nL de IIB+H₂ T3 - T2
- CSA / NRTL: Class I, Div. 1, Gas groups B, C, D with Type Y-Purge, Class I, Div. 2, Gas groups B, C, D, Temperature code T3 - T2
Power Requirement
- 100-240 Vac, 50 to 60 Hz
Power Consumption
- Typical, varies with installed options
- 1,200 W Startup
- 900 W steady-state Operation
Tube Connection
- Supply: 3/8” Tube, minimum
Pressure
- Inlet: Set Point 69 kPa (10 psig) to 1,034 kPa (150 psig) Max.
- Supply: 414 kPa (60 psig)
Flow Rate
- Steady State: 127 to 147 L/min (4.5 to 5.2 ft³/min) at 20° C Y-purge Types
Material of Construction
- Oven liner: Stainless Steel
Ambient Temperature
- Oven: 30° to 180° C
Resolution
- Setpoint: 1° C, 0.01 psig (Gas Control)
Stability
- Ambient Range: ± 1.0° C
- Temperature (Steady ambient): ± 0.1° C
Accuracy
- 0 to 50 psig: 1.7 %
- 50 to 100 psig: 2.7 %
Repeatability
- ± 0.1 psig
- A designation was assigned to the ABB On-line sulfur in fuel gas chromatograph (PGC2007) - ASTM Method D7041-04.
- The PGC2007 has now been updated to the PGC5000 Series PGC5007B Total Sulfur Analyzer.
- This method provides industry the only on-line method with precision and bias data that is 2-5 times better than all other sulfur methods on the books.
- This is the only method to date that meets the requirements for low range (0-15 ppm) diesel compliance measurements.
Documentation
Brochure
- Bulletin - PGC5000 Series - Factory notice - Safety reminder pdf 398 KB
- Bulletin - PGC5000 Series Version 4.2.1 Release is now available pdf 151 KB
- Bulletin - Time Synchronization in PGC5000 Data System pdf 218 KB
- Bulletin - VistaSTAR with PGC5000 Version 4 pdf 236 KB
- Course description - Z234e Process Gas Chromatography | External version Summary: No summary available pdf 49 KB
- Windows Server 2003 OS - Support ended on July 14, 2015 pdf 210 KB
- BB PGC5000 Series technology review | Solutions from the process sample tap to measurement delivery Summary: High Resolution pdf 3.1 MB
- Refinery Flare Monitor | Analytical Solutions for 40CFR60 subpart Ja pdf 4.7 MB
- Refinery Flare Monitor | Analytical Solutions for 40CFR60 subpart Ja | Press-ready file pdf 10.6 MB
Catalog
Certificate
- Certificate of conformity - PGC5000 Series IECEx LCI 09 00 10X Issue 04 pdf 2.6 MB
- PGC5000 Series - PGC5007 - EMC Summary pdf 266 KB
- CE - Declaration of Conformity. PGC5000, PGC5000A, PGC5000B, PGC5000C, PGC5007, PGC5009 pdf 131 KB
- CE PGC5000 Series pdf 82 KB
- Declaration of Conformity - PGC5000 pdf 171 KB
- PGC5000 Series LCIE 09ATEX3006 Version 09 pdf 649 KB
- PGC5000 Series LCIE 09ATEX3006X Zone 1 pdf 2.1 MB