Brand: Grabner InstrumentsThe MINISCAN IR VISION is a high speed, compact and robust FTIR fuel analyzer for the comprehensive and automatic measurement of gasoline, jet and diesel fuels.
The analyzer is configured to measure more than 100 fuel parameters and components for fuel blending, for quality inspection and to check compliance with fuel specifications directly at the point of sale.
100+ Preconfigured Parameters Measured
Compound Analysis is performed according to the international standards ASTM D5845 for oxygenates, ASTM D6277 and EN 238 for Benzene and EN 14078 for Biodiesel blends.
A built-in temperature regulated density meter allows for exact determination and stability of fuel density.
Octane and Cetane Number, Distillation, Vapor Pressure and other fuel properties are automatically determined from the full IR spectrum using Partial Least Square (PLS) analysis and advanced chemometric models following ASTM E1655.
Several thousand data points are used to achieve best prediction accuracy.
Data Based on Real Samples Collected and Analyzed by SGS®!
For predicted properties, MINISCAN IR VISION operates calibration data, that has to adhere to the strictest quality standards.
As a unique feature the instrument is equipped with a database of samples that have been collected and analyzed by SGS!
High Quality Portable Technology
MINISCAN IR Vision is unmatched in its class of portable fuel analyzers.
The thermoelectric temperature regulation of the instrument´s filling system, measuring cells and integrated density meter maximizes accuracy in measuring volume and mass percent of fuel compounds.
Robust for Field Use
Durability makes the instrument ideal for the challenges encountered during field or mobile testing. The MINISCAN IR Vision incorporates Grabner´s proven, robust and bubble free metal filling system.
The instrument is protected by the shock and vibration tested Vision platform housing.
The double interferometer is mounted with a self aligning mirror system, that allows automatic correction of intensity shifts after a rough drive over a bumpy road.
A robust, 10” full color industrial touchscreen guarantees the highest visibility and ease of use even under rough environmental conditions. For field use, the instrument can be run with a 12V car adapter.
Models
Dedicated models for Gasoline or Diesel analysis are available. The PRO model performs analysis of Gasoline, Diesel and Jet Fuel. Dedicated models can be upgraded to a professional model.
Features- Portable Fuel Analyzer for Gasoline, Diesel, Jet Fuel and Biofuel Blends
- Contamination Detection: Automatic Distance Analysis between Spectra
- 10” Industrial Full Color Touchscreen
- Smart 2+1 Cell Design
- Laser Controlled Spectrometer
- Full Spectrum PLS Analysis Using Superior Processing Power
- Beam Splitter: Ge-Coated KBR
- Bubble Free and Robust Metal Filling System
- Thermoelectric Temperature Regulation of Filler, Density Meter and Cells
- Remote Access. Anywhere. Anytime
Specifications
Brand
Models Available
- MINISCAN IR VISION Gasoline
- MINISCAN IR VISION Diesel
- MINISCAN IR VISION Professional (Gasoline, Diesel and Jet Fuel)
Gasoline Parameters - Properties & Range
- RON: 70 - 110
- MON: 65 - 105
- AKI: 67 - 107
- Distillation: IBP, T10, T50, T90, FBP
- Evaporation: E70, E100, E150 (°C), E200, E300 (°F)
- RVP & DVPE: 40 - 105 kPa
- Density: 0 - 3 g/cm³
- Precision (Density): R s.d. = ± 0.0005 g/cm³
- Driveabilty Index
- VOC Emissions Calculator
- Vapor Lock Index (VLI)
- Unlimited User-Defined Parameters
Gasoline Parameters - Aromatics & Range
- Benzene: 0 - 10 m %
- Toluene: 0 - 20 m %
- m-Xylene: 0 - 20 m %
- o-Xylene: 0 - 20 m %
- p-Xylene: 0 - 20 m %
- Ethyl Benzene: 0 - 20 m %
- Propyl Benzene: 0 - 20 m %
- Mesitylene: 0 - 20 m %
- Durene: 0 - 20 m %
- Naphthalene: 0 - 10 m %
- Pseudocumene: 0 - 20 m %
- 2-/3-/4-Ethyl Toluene: 0 - 20 m %
- Other Aromatics: 0 - 20 m %
Gasoline Parameters - Oxygenates & Range
- MTBE: 0 - 20 m %
- TAME: 0 - 20 m %
- DIPE: 0 - 20 m %
- Ethanol: 0 - 40 m %
- Iso-Propanol: 0 - 20 m %
- 2-Butanol: 0 - 25 m %
- tert-Butanol: 0 - 25 m %
- Sec-Butylacetate: 0 - 10 m %
- Iso-Butylacetate: 0 - 10 m %
- Dimethylcarbonate: 0 - 10 m %
- Dimethoxymethane: 0 - 10 m %
- Acetone: 0 - 25 m %
- Other Oxygenates: 0 - 20 m %
- ETBE: 0 - 20 m %
- Methanol: 0 - 15 m %
Gasoline Parameters - Octane Boosters & Range
- MMT/CMT (mg/l): 0 - 10,000
- Manganese (MMT): 0 - 2500
- DCPD: 0 - 15 m %
- Nitromethane: 0 - 9 m %
Gasoline Parameters - Others
- Cyclohexane: 0 - 100 m %
Gasoline Parameters - Aniline & Range
- Aniline: 0 - 5 m %
- N-Me-Aniline: 0 - 5 m %
- N, N-Dimethylaniline: 0 - 5 m %
- m-Methylaniline: 0 - 5 m %
- o-Methylaniline: 0 - 5 m %
- p-Methylaniline: 0 - 5 m %
Gasoline Parameters - Top Parameters & Range
- Total Oxygen: 0 - 12 m %
- Total Aromatics: 0 - 80 m %
- Total Olefins: 0 - 80 m %
- Di-Olefins: 0 - 20 m %
- Total Aniline: 0 - 5 m %
- Total Esters: 0 - 5 m %
Diesel Parameters - Properties & Range
- Cetane Number: 20 - 80
- Cetane Index: 20 - 80
- Kinematic Viscosity @ 40° C: 0 - 10 mm²/s
- Dynamic Viscosity @ 40° C: 0 - 10 mPas
- CFPP: -50° C to +20° C
- Distillation / Recovery: IBP, T10/50/65/85/90/95, FBP, R250, R350
- Density: 0 - 3 g/cm³ (r s.d. = ± 0.0005 g/cm³)
Diesel Parameters - Components & Range
- Total Aromatics: 0 - 80 m %
- Poly Nuclear Aromatics: 0 - 50 m %
- Cetane Improver: EHN, IPN: 0 - 10,000 ppm
- Biodiesel (FAME, FAEE): 0 - 40 v %
Jet Fuel Parameters - Properties & Range
- Flashpoint: -20° C to +100° C
- Freezing Point: -80° C to +20° C
- Kinematic Viscosity @ -20° C: 0 - 10 mm²/s
- Distillation: IBP, T10/ 50/90/95, FBP, E10/50, R200
- Smoke Point: 0 - 1000 mm
- Total Aromatics: 0 - 40 m %
- Naphtalenes: 0 - 5 m %
- MSEP: 60 - 100 %
- Density: 0 - 3 g/cm³ (rs.d.= 0.0005 g/cm³)
Jet Fuel Parameters - Components & Range
- Biodiesel (FAME): 0 - 0.12 m %
Standards
- ASTM D1655, D5845, D6277, D7777, D7806, E1655, EN 238, EN 14078, ISO 15212
Correlation to
- ASTM D56/D3828, D86, D323, D445, D1319, D1322, D1655, D1840, D2386/D7153, D3948, D5191, D6371, D6378, D6379, D613, D2699, D2700, ISO 3104, 3405, 5163, 5164, 5165, EN 116, EN 13016
Spectrometer
- Temperature and Laser Regulated, 2+1 Cell-FTIR
Density Measurement
- Temperature Regulated Oscillating U-Tube Cell
Warm Up / Scanning Time
- < 30 s / 80 s (Multiple Scans)
Units of Measurement
- v %, m %
Touch-Screen
- 10” Full Color Touch Screen
Interfaces
- 2 x USB, 2 x LAN
Power Supply
- 100-264 Vac, 47-63 Hz, 130 W (Field Application: DC Adaptor for 12 V Vehicle Battery)
Dimensions / Weight
- 293 x 390 x 280 mm (10.5” x 15.4” x 11”)/ 12 kg (26 lb)