Yodify Product Library
Instantly add the FC500G Free Available Chlorine Analyzer and 50,000+ more product series to your store or catalog
Book Your Demo and See How
or create your store
FC500G Free Available Chlorine Analyzer
Brand: Yokogawa Electric CorporationFC500G non-reagent free available chlorine analyzer adopts the polarographic method, using a rotating electrode to continuously measure online the concentration of free available chlorine.
The FC500G is intended for measurements in the water distribution network, at pumping stations, water delivery points, schools, and public facilities.
Features- Measuring range can be selected from 0 to 1 / 0 to 2 / 0 to 3 mg/l and 0 to 5 / 0 to 10 mg/l.
- Rotating electrode/ceramic bead cleaning has been improved to making cleaning automatic and continuous.
- Easy-to-maintain (-clean) cell construction.
Specifications
Air Consumption
- About 5 LPM
Ambient Temperature Range
- -5° to 50° C (23° to 122° F)
Approvals & Certifications
- EMC: EN 55011 Class A, Group 1
Compensated Temperature
- 5° to 35° C (41° to 95° F) - Sample Solution Temperature
Consumption
- Approximately 15 VA (100 or 110 V) - Power Consumption
- Approximately 20 VA (200 or 220 V) - Power Consumption
Enclosure Finish
- Baked Polyurethane Resin Coating
- Munsell 5Y7/1
Flow Rate
- 0.026 to 0.660 gpm (0.1 to 2.5 LPM) at Cell Inlet - Sample Solution Flow Rate
Frequency
- 50/60 Hz
Indication
- 3-1/2-Digit LCD
Linearity
- ± 5 % of Full Scale
Load Resistance
- Maximum
Measuring Range
- 0 to 10 mg/L - Air Purging, Sample Solution SS (Suspended Solid)
- 100 to 300 µS/cm - Sample Solution Conductivity
- pH: 6.5 to 7.5 - Operating PH Range of Sample Solution
Mounting
- 2” (50.8 mm) Pipe Mounting
- Wall Mounting
Output
- 4-20 mA DC (Load Resistance: Max.
Power Supply
- 100 Vac ± 10 %, 50/60 Hz
- 110 Vac ± 10 %, 50/60 Hz
- 200 Vac ± 10 %, 50/60 Hz
- 220 Vac ± 10 %, 50/60 Hz
Power Supply Effect
- ± 0.5 % of Full Scale / 10 % of Rated Voltage - Power Voltage Fluctuation Influence
Pressure Range
- psi: 0.145 to 21.76 (kPa: 1 to 150, bar: 0.01 to 1.5) with Valve - Sample Solution Pressure
- psi: 0.145 to 4.351 (kPa: 1 to 30, bar: 0.01 to 0.3) without Valve - Sample Solution Pressure
Repeatability
- 2 % of Full Scale - Reproducibility
Response Time
- Approximately 2 minutes
RTD Type
- Pt1000
Span & Zero Effect
- Span Drift: Max.
- Zero Drift: within ± 1 % of Full Scale/Month (Stabilty)
Supply Pressure
- psi: 7.25 to 20.31 (kPa: 50 to 140, bar: 0.5 to 1.4) - Supply Air Pressure
Temperature Effect
- ± 0.5 % of Full Scale for 10° C (18° F) Change - Ambient Temperature Influence
- ± 3 % of Full Scale/5° to 35° C (41° to 95° F) - Sample Solution Temperature Compensation Error
Temperature Range
- 0° to 40° C (32° to 104° F) - Sample Solution Temperature
Wetted Materials
- Acrylic Resin (Molding) - Measuring Tank
- Flexible Polyethylene (PE) - Piping
- Flexible PVC - Piping