Yodify Product Library
Instantly add the DM-501-ClO₂ Chlorine Dioxide Sensor and 50,000+ more product series to your store or catalog  
Book Your Demo and See How
or create your store

DM-501-ClO₂ Chlorine Dioxide Sensor

Brand: Detcon Inc

Detcon Model DM-501IS-ClO₂ sensor is designed to detect and monitor CLO2 in air over the range 0-1 ppm. 

The sensor assembly consists of an electrochemical fuel cell mounted in an Intrinsically Safe stainless steel and ABS plastic housing that includes a splashguard with integral cal port, a signal conditioning control transmitter circuit, a base connector board and an explosion proof junction box. 

The Intrinsically Safe sensor housing design allows direct contact of the target gas to the electrochemical sensor, thus maximizing response time, detectability and repeatability. 

And because the Intrinsically Safe housing design does not use a stainless steel flame arrestor, it eliminates inaccurate readings due to target gas absorption.

1 - 2 of 2 Part Numbers
965-777780-001
965-777780-001
  • Junction Box - Aluminum
965-77778B-001
965-77778B-001
  • Junction Box - Stainless Steel
1 - 2 of 2 Part Numbers

Specifications

Brand

Approvals & Certifications

  • CSA, UL Class I, Div. 1, Groups B, C, D
  • ISO 9001: 2008
  • NRTL
  • UL Class I, Div. 1, Groups B, C, D

Consumption

  • 1.6 W at 24 Vdc

Display

  • Backlit LCD Display

Electrical Connectors

  • Junction Box

Enclosure, Body Material

  • ABS
  • Stainless Steel

Environmental Protection

  • Explosion-proof
  • Intrinsically Safe
  • IP66
  • NEMA 4X

Humidity

  • 10 to 95 % Relative Humidity, Non-Condensing

Indication

  • LED Indicators for Fault & Cal Status

Input

  • 11.5-28 Vdc

Materials of Construction

  • Aluminum - Junction Box
  • Stainless Steel - Junction Box

Measuring Range

  • 0 to 1 ppm Chlorine Dioxide

NPT Connection

  • ¾” NPT

Operating Temperature

  • -20° to 40° C (-4° to 104° F)

Output

  • 4-20 mA DC

Repeatability

  • ± 2 % of Full Scale

Response Time

  • T90 < 120 seconds

Sensor, Probe Type

  • Electrochemical Fuel Cell

Span Effect

  • < 1 % Signal Loss Per Month

Instantly populate your store or catalog 
CREATE YOUR SITE
or book a demo with the Yodify Team
show more