PGC5000B Smart Oven Process Gas Chromatograph
Brand: ABB GroupABB process gas chromatographs perform real-time analysis of the chemical composition of a process sample or stream - crucial for on-spec processes and productivity.
PGC5000B Smart Ovens target simple applications with a fixed set of features.
The ABB PGC5000 Series process gas chromatograph is based on over 50 years of innovation in on-line process gas chromatographs.
It simplifies gas chromatography and increases reliability. It provides everything you need to know about your GC – at your fingertips.
The PGC5000 Series has been designed for distributed control across multiple analytical Smart Ovens for limitless application approaches.
There are two ovens, PGC5000B and PGC5000C, suitably equipped for the type of application engineered.
PGC5000B Smart OvenTM technology supports applications that require a single detector with up to three analytical valves.
Advanced pressure, temperature and stream control software executes analytical methods required for analyses.
A single PGC5000A Controller can support up to four PGC5000B Smart OvensTM reducing space and utility requirements while increasing applications flexibility.
Features- Correctly sized and designed for simple configurations and maximum reliability
- Designed for mission critical measurements
- Standard detector configuration – sTCD, mTCD, FID and FPD capability
- Smaller footprint allows for shelter size and utility reduction
Specifications
Brand
Environmental Protection
- IP54 (NEMA 3 Equivalent)
Ambient Temperature
- Oven: 30° to 180° C (Settings and display in °C only)
- 0° to 50° C (32° to 122° F)
Humidity
- 95 % Relative Humidity, Non-Condensing
Mounting
- Wall: 33 mm (1.3 in.) from wall with brackets
EMI / RFI
- Conform to ClassA industrial environment
Certifications & Approvals
- ATEX / IEC / CN / KO: Zone 1: CE 0344, II2G, Ex de py IIB+H₂ T4 - T2, Zone 2: CE, II3G Ex de nA nL IIB+H₂ T4 - T2 Ex de px IIB+H₂ T4 - T2 (Optional), with X-purge Power interlock
- CSA / NRTL: Class I, Div. 1, gas groups B, C, D with Type Y-purge,
Power Requirement
- 100-240 Vac, 50/60 Hz
Power Consumption
- 900 W steady-state Operation
- 1,200 W Startup
- Typical, varies with installed options
Pressure
- Inlet: Max.: 1,034 kPa (150 psig)
- Setpoint: 69 kPa (10 psig)
- Supply: 551.6 kPa (80 psig)
Flow Rate
- Steady State purge: 127 to 147 L/min (4.5 to 5.2 ft³/min) at 20° C, Y-purge Types
Material of Construction
- Tube: 316 Stainless Steel
- Oven liner: Stainless Steel
Resolution
- Setpoint: 1° C
Stability
- Ambient Range: ± 1.0° C
- Steady ambient: ± 0.1° C
Accuracy
- 0-100 psig: 2 % (Gas Control)
Repeatability
- ± 0.05 psig (Gas Control)
Media
- H₂, He, N₂, Air, Ar
- No Liquids, corrosives, combustibles, O₂
Tube Connection
- 1/16”, 1/8”, 1/4” Gyrolok/Swagelok Fitting
Documentation
Brochure
- Bulletin - PGC5000 Series - Factory notice - Safety reminder pdf 398 KB
- Bulletin - PGC5000 Series Version 4.2.1 Release is now available pdf 151 KB
- Bulletin - Time Synchronization in PGC5000 Data System pdf 218 KB
- Bulletin - VistaSTAR with PGC5000 Version 4 pdf 236 KB
- Windows Server 2003 OS - Support ended on July 14, 2015 pdf 210 KB
- Z234e Process Gas Chromatography | External version pdf 49 KB
Catalog
- Data Sheet - PGC5000 Integrated controller | A4 pdf 7.0 MB
- Data Sheet - PGC5000 Integrated controller | US pdf 6.8 MB
- Data sheet - PGC5000 Temperature Programmed Gas Chromatograph pdf 6.3 MB
- Data sheet - PGC5000B Smart Oven pdf 211 KB
- Data sheet - Zone 2 Certification by design PGC5000 Series pdf 243 KB
- PGC5000A Generation 2 Master controller pdf 1.5 MB