MU5 Free Range Universal Temperature Converter
Brand: Yokogawa Electric CorporationThe MU5 is a plug-in type universal temperature converter that converts input signal (thermocouple, RTD or mV signal) into isolated DC current or DC voltage signals.
Features- Selection of input type(thermocouple, RTD or mV signal), I/O range setting, burnout setting, output adjustment, I/O monitoring, and loop back test can be made using the optional Parameter Setting Tool (VJ77) or Handy Terminal (JHT200).
- The operation indicating lamp shows the operation status, abnormalities in a setting etc.
- Output adjustment, wiring resistance correction, and ON/OFF of RJC can be made using the switches on the front panel of the MU5 without a setting tool such as Handy Terminal.
- For the Fahrenheit display, specify the option “/DF”.
Specifications
Accuracy
- ± 0.05 % - Output Accuracy
- ± 0.05 % of Span or ± 0.05° C (± 0.09° F) whichever is greater - for Pt50 (JIS’81)
- ± 0.1 % of Span
- ± 0.1 % of Span or ± 1° C (± 1.8° F) whichever is greater - Typical K, E & T: less than -200° C (-328° F), Typical B: 400° C (752° F) to less than 600° C (1,112° F), Typical E & J: more than 750° C (1,382° F), Typical N: more than 1,200° C (2,192° F)
- ± 0.1 % of Span or ± 2° C (± 3.6° F) whichever is greater - Typical N: less than -200° C (-328° F)
Burn-Out
- 60 seconds - Max.
- Downscale
- Off
- Upscale
Consumption
- 1.6 W at 110 Vdc - Power Consumption
- 1.7 W at 24 Vdc - Power Consumption
- 3.5 VA at 100 Vac - Power Consumption
- 4.9 VA at 200 Vac - Power Consumption
Current
- 0.7 mA Approximately - RTD Detective Current
Electrical Connectors
- M3.5 Screw Terminal
Humidity
- 5 to 90 % Relative Humidity, Non-Condensing - Operating Humidity
Input
- ± 4 Vdc - Max.
- -500 to 500 mVdc - Measuring Range
Insulation Resistance
- 100 megaohms at 500 Vdc between Input, Output, Power Supply & Grounding Terminals Mutually
Materials of Construction
- ABS Resin (Black), UL 94 V-0 - Main Unit
- ABS Resin + Polycarbonate Resin (Black), UL 94 V-0 - Main Unit
- Modified Polyphenylene Oxide Resin, Including Glass Fiber (Black), UL 94 V-1 - Socket
- PBT Resin, Including Glass Fiber (Black), UL 94 V-0 - Main Unit
Measuring Span
- 10° C (50° F) or more (RTD)
- 3 mV or more (Thermocouple, mV Signal)
Mounting
- DIN Rail Mounting
- Plug-In Typical
- Wall Mounted
Operating Temperature
- 0° to 50° C (32° to 122° F)
Operating Voltages
- 12-48 V
- 85-264 V
Output
- ± 10 Vdc, Span is 0.1 V or more
- ± 100 mVdc, Span is 10 mV or more
- 0-20 mA DC, Span is 5 mA or more
- 0-5 mA DC - Span is 1 mA or more
- 15 V/Max.
Power Supply
- 100-240 Vac/dc (-15, 20 %), 50/60 Hz - Power Supply Input & Rated Voltage
- 15-40 Vdc (± 20 %) - Power Supply Input & Rated Voltage
Power Supply Effect
- ± 0.1 % of Span or Less for the Fluctuation within the Operating Range of Each Power Supply Voltage Specification - Effect of Power Supply Voltage Fluctuations
Resistance
- ± 0.2° C (0.36° F) or Less for a Change of 10 ohms / Wire - Effect of Leadwire Resistance for RTD
- ± 15 µV or Less for a Change of 100 ohms - Leadwire Resistance Change Effect for Thermocouple
- 1 Megaohm During Power On - Input Resistance
- 10 kiloohms During Power Off (Thermocouple, mV Signal) - Input Resistance
- 10 kiloohms or more for ± 10 Vdc - Allowable Load Resistance for Voltage Output
- 2 kiloohms or more for ± 5 Vdc - Allowable Load Resistance for Voltage Output
- 250 kiloohms or more for ± 100 mVdc - Allowable Load Resistance for Voltage Output
- 500 ohms or Less - Input External Resistance
Response Time
- 150 milliseconds, 63 % Response (10 to 90 %)
RTD Type
- JPt100 (JIS’89)
- Pt100 (ITS-90: JIS’97 & IPTS-68: JIS’89)
- Pt50 (JIS’81)
Sensor, Probe Type
- B Thermocouple (ITS-90: JIS’97)
- E Thermocouple (ITS-90: JIS’97)
- J Thermocouple (ITS-90: JIS’97)
- K Thermocouple (ITS-90: JIS’97)
- N Thermocouple (ITS-90: JIS’97)
- RTD
- R Thermocouple (ITS-90: JIS’97)
- S Thermocouple (ITS-90: JIS’97)
- Thermocouples
- T Thermocouple (ITS-90: JIS’97)
- W3 Thermocouple
- W5 Thermocouple
Span & Zero Adjustments
- ± 1 % of Span or More - Input Adjustment
- ± 5 % of Span or More - Output Adjustment
Supply Voltage
- 2,000 Vac for 1 Minute between Input, Output, Power Supply & Grounding Terminals Mutually - Withstand Voltage
Temperature Effect
- ± 0.15 % of Span or Less for a Temperature Change of 10° C (50° F) - Ambient Temperature Effect
Temperature Range
- -200° to 850° C (-328° to 1,562° F) - RTD Measuring Range
- -270° to 2,300° C (-454° to 4,172° F) - Thermocouple Measuring Range