Yodify Product Library
Instantly add the GM90MB Module Base and 50,000+ more product series to your store or catalog  
Book Your Demo and See How
or create your store

The Data Acquisition System GM is a data logger that excels in versatility and expandability. The main unit includes on-board memory for data acquisition and also supports SD cards for external storage.

The system consists of a Data Acquisition Module (GM10), Power Supply Module (GM90PS), and Module Base (GM90MB), which is used to mount a variety of modules.
Features
Scalability:
  • Up to 420 channels per system
  • Plug and lock modules
Ease of Use:
  • Web-based configuration
  • Live web-based data viewing
Mobile Connectivity:
  • Bluetooth
  • Mobile application
Reliability:
  • Secure data storage
  • High accuracy measurement
  • Automatic backfill function (GA10 data logging software)
Noise Tolerance:
  • Electromagnetic relay module

Specifications

Ambient Temperature Range

  • -20° to 60° C (-4° to 140° F)

Approvals & Certifications

  • CE
  • CSA 22.2 No61010-1, CSA 22.2 No.61010-2-030-12
  • EMC: EN 301 489-1, EN 301 489-17, EN 61326-1, EN 61326-1 Class A Table 2, EN 61000-3-2, EN 61000-3-3, EN 55011 Class A Group 1
  • KC
  • LVD: EN 61010-1, EN 61010-2-030
  • NRTL
  • RCM: AS/NZS4268, AS/NZS2772.2
  • UL 61010-1, UL 61010-2-030 (CSA NRTL/C)
  • WEEE Directive Support

Approvals & Certifications

  • R & TTE Directive - EN 61010-1, EN 61010-2-030, EN 62311

Communication

  • Bluetooth
  • DARWIN
  • Ethernet - 10Base-T/100Base-TX Port
  • EtherNet/IP
  • Modbus RTU
  • Modbus TCP/IP
  • RS-422, RS-485
  • Serial Communication Port
  • SSL
  • USB 2.0
  • Wireless

Display

  • 7-Segment LED

Electrical Connectors

  • M3 Screw Terminal

Frequency

  • 50/60 Hz ± 1 %

Humidity

  • 20 to 85 % Relative Humidity, Non-Condensing - Ambient Humidity
  • 55 ± 10 % Relative Humidity, 23 ± 2° C (73.4 ± 3.6° F) - General Operating Condition

Input

  • 20 mV - DCV
  • -200 mV to 50 V - DCV
  • 6 V (1-5 V) - DCV

Measurement Accuracy

  • DCV: ± (0.01 % of Reading + 2 mV) - 6 V (1-5 V)
  • DCV: ± (0.01 % of Reading + 5 μV) - 20 mV
  • ± [0.02 % of Reading + 0.16° C (0.288° F)] - Pt100 (High Resolution)
  • ± [0.02 % of Reading + 0.2° C (0.36° F)] - Pt100
  • ± 0.2° C (± 0.36° F), However, -200° to 0° C (-328° to 0° F): ± [0.15 % of Reading + 0.2° C (0.36° F)] - K Thermocouple [-200° to 500° C (-328° to 932° F)]
  • ± 0.2° C (± 0.36° F), However, -200° to 0° C (-328° to 32° F): ± [0.10 % of Reading + 0.2° C (0.36° F)] - T Thermocouple
  • ± 1.1° C (± 1.98° F) - R Thermocouple

Mounting

  • Clamp Terminal
  • Desktop
  • DIN Rail
  • Wall Mounted

Operating Temperature

  • -20° to 60° C (-4° to 140° F)

Resistance

  • 1 Kiloohm or Less (ON) - Contact DI
  • 100 kiloohms or more (OFF) - Contact DI

RFI/EMI Effect

  • 400 A/m or Less (DC & 50/60 Hz) - Magnetic Field

RTD Type

  • Cu10
  • Cu100
  • Cu25
  • Cu50
  • Cu53
  • JPt100
  • Ni100
  • Ni120
  • Pt100
  • Pt200
  • Pt25
  • Pt46
  • Pt50

Sensor, Probe Type

  • B Thermocouple
  • E Thermocouple
  • J Thermocouple
  • K Thermocouple
  • L Thermocouple
  • N Thermocouple
  • RTD
  • R Thermocouple
  • S Thermocouple
  • T Thermocouple
  • U Thermocouple
  • W Thermocouple

Shock

  • When Off, 500 m/sec² or Less, Approximately 10 milliseconds, 3 times in 6 Directions (± X, ± Y, ± Z)
  • When On, 98 m/sec² or Less, 11 milliseconds, 3 times in 6 Directions (± X, ± Y, ± Z)

Temperature Range

  • -200° to 850° C (-328° to 1,562° F) - RTD
  • -270° to 2,320° C (-454° to 4,208° F) - Thermocouples

Update Time

  • 30 minutes - Warmup Time

Vibration

  • 5 ≤ F < 8.4 Hz Amplitude 3.5 mm Peak
  • 8.4 ≤ F ≤ 160 Hz Acceleration 9.8 m/sec²
Instantly populate your store or catalog 
CREATE YOUR SITE
or book a demo with the Yodify Team
show more