Yodify Product Library
Instantly add the ECP 20-2 Peltier Gas Cooler and 50,000+ more product series to your store or catalog  
Book Your Demo and See How
or create your store

The M&C gas cooler ECP 20-2 is used in gas analysis to lower the dew point of humid gas to avoid condensation in the analyzer. 

An extremely stable gas dew point avoids water vapour cross-sensitivity and volumetric errors.
Features
  • Compact, self-controlling and maintenance-free
  • With Jet-Stream heat exchanger available in 3 standard materials
  • Gas flow rate 1 stream 500 l/hr
  • Outlet dew point adjustable from +2° to 15° C
  • Dew point stability ±0.1° C
  • Status-Alarm contact standard
  • Small dimensions
  • Light weight
  • Optimum reliability

1 - 3 of 3 Part Numbers
01K8100
01K8100
  • Heat Exchanger - Duran Glass
01K8120
01K8120
  • Heat Exchanger - Stainless Steel 316Ti
01K8140
01K8140
  • Heat Exchanger - PVDF
1 - 3 of 3 Part Numbers

Specifications

Ambient Temperature Range

  • 5° to 45° C (41° to 113° F)

Approvals & Certifications

  • Electrical Standard: EN 61010

BSP Connection

  • G 1/4”i
  • G 3/8”i

Capacity

  • Total Cooling Capacity at 25° C (77° F) Ambient: 22.22 W (80 KJ/h)

Consumption

  • Power Consumption: 250 VA

Electrical Connectors

  • 2 x PG 11 Cable Gland
  • Terminals 2.5 mm²

Electrical Rating

  • Status Alarm 1 Changeover Contact, Potential Free: 2 A at 250 V, 500 VA, 50 W Alarm Point: ∆ ± 3° C to TSOLL

Enclosure Finish

  • Black

Environmental Protection

  • IP54 EN 60529

Flow Rate

  • Gas Stream/Gas Flow Rate: 1 X, Max.

Materials of Construction

  • Material of Heat Exchanger: 316Ti Stainless Steel
  • Material of Heat Exchanger: Duran Glass
  • Material of Heat Exchanger: PVDF

Mounting

  • Wall Mounting

Power Supply

  • 115 V, 60 Hz Switchable
  • Main Power Connection: 230 V, 50 Hz

Pressure Range

  • Gas Pressure, Max.

Response Time

  • Ready for Working: < 20 minutes

Stability

  • Dew Point Stability: ± 0.1° C (± 0.18° F)

Storage Temperature

  • -20° to 60° C (-4° to 140° F)

Temperature Range

  • Sample Outlet Dew Point: 2° to 15° C (35.6° to 59° F), Factory Setting 5° C (41° F)

Tube Connection

  • Condensate Connection: Ø 6 mm O.D. Tube G 3/8”i
  • Sample Gas Connection: Ø 12 mm O.D. Tube G 1/4”i

Instantly populate your store or catalog 
CREATE YOUR SITE
or book a demo with the Yodify Team
show more