OxyTrans-EX Explosion-Proof Oxygen Transmitter
Brand: Edgetech Instruments4-20 mA Loop Powered Two Wire Oxygen Transmitter
- The Edgetech Instruments Model OxyTrans is an easy to use, accurate and economical way to measure Oxygen at trace and % levels. A simple display and one point calibration allows for a quick start of the instrument.
Measuring Range
- The unit will offer ranges between 0-100ppm, 0-1,000ppm, 0-10,000ppm of oxygen. 0-1% to 0-25% O2.
Sensor
- The OxyTrans uses a special fuel cell to measure the oxygen concentration. The sensor meets the industrial requirements for accuracy, sensitivity, easy to use and operating life.
Calibration
- The calibration of the instrument for trace oxygen measurements in gas should be done with a known calibrated gas. The concentration can be chosen freely within the measuring range.
Flow-Through Measuring Cell in Stainless Steel
- The measuring cell is modular and made of stainless steel. A defective measuring cell can be repaired by replacing the defective part only, rather than the complete unit.
- Products
- About Us
- News & Events
- Customer
- A 2 wire explosion-proof in-line oxygen transmitter
- II 1/2 G Ex ia IIC T4 (ATEX approved)
- 4-20mA output
- Display: 6 digit alphanumeric with bar graph
- 5 ranges offered: 0-100 ppm, 0-1,000 ppm, 0-10,000 ppm, 0-1%, 0-25%
- Oxygen sensor: a micro fuel cell
- Easy, accessible potentiometer
Specifications
Brand
Measuring Ranges
- O₂ ppm Range: 0-100 ppm
- 0-1,000 ppm
- 0-10,000 ppm
- O₂ % Range: 0-1 %, 0-25 %
- Range must be specified at time of ordering
Calibration
- With calibrated gas, air
Accuracy
- ± 2 % FSD T = constant
- ± 5 % FSD 0 > T > 50° C
Resolution
- 0.1 ppm at 0-100 ppm
- 1 ppm at 0-1,000, 0-10,000 ppm
- 0.1%
Response Time
- 90 % FSD at 25° C
- 0-10 ppm < 45 s
- 0-100 ppm < 20 s
- 0-1,000 ppm < 10 s
Operating Temperature
- 0° to 50° C
Pressure
- 0.1 - 1 bar
Signal Output
- 4-20 mA/dc
Alarm Value
- 3.8 mA (Standard)
- 23 mA (Optional)
Voltage
- 10 - 35 Vdc
Load Range
- 100 to 750 ohm
- Typical 470 ohm
Display
- 6 Digits, alphanumeric with bar graph
Oxygen Sensor
- Micro-Fuel Cell
Housing
- IP65
Dimensions
- 120 mm x 160 mm x 65 mm (L x H x W)
- 4.7” x 6.3” x 2.5”
Weight
- 1.2 kg (206 lb)
Options
- Different sensors available
- 5 different ranges
- 316 Stainless Steel cell Bl
Model Types
- OxyT: Standard
- OxyT-H: Works with the Hart Protocol
- OxyT-RA: Right angle display
- OxyT-SS: Panel mount sampling system, Panel size: 38.1 cm (15”) square, Weight: 3.2 kg (7 lb)
- OxyT-PSS: Portable sampling system with power supply and pump option, Weight: 6.8 kg (15 lbs)
- OxyTrans-EX: Explosion Proof, 2-Wire loop powered transmitter (ATEX approved)
- Air Separators