OxyTrans-EX Explosion-Proof Oxygen Transmitter

Brand: Edgetech Instruments

4-20 mA Loop Powered Two Wire Oxygen Transmitter

  • The Edgetech Instruments Model OxyTrans is an easy to use, accurate and economical way to measure Oxygen at trace and % levels. A simple display and one point calibration allows for a quick start of the instrument.
Measuring Range

  • The unit will offer ranges between 0-100ppm, 0-1,000ppm, 0-10,000ppm of oxygen. 0-1% to 0-25% O2.
Sensor

  • The OxyTrans uses a special fuel cell to measure the oxygen concentration. The sensor meets the industrial requirements for accuracy, sensitivity, easy to use and operating life.
Calibration

  • The calibration of the instrument for trace oxygen measurements in gas should be done with a known calibrated gas. The concentration can be chosen freely within the measuring range.

Flow-Through Measuring Cell in Stainless Steel

  • The measuring cell is modular and made of stainless steel. A defective measuring cell can be repaired by replacing the defective part only, rather than the complete unit.
    • Products
    • About Us
    • News & Events
    • Customer
Features
  • A 2 wire explosion-proof in-line oxygen transmitter
  • II 1/2 G Ex ia IIC T4 (ATEX approved)
  • 4-20mA output
  • Display: 6 digit alphanumeric with bar graph
  • 5 ranges offered: 0-100 ppm, 0-1,000 ppm, 0-10,000 ppm, 0-1%, 0-25%
  • Oxygen sensor: a micro fuel cell
  • Easy, accessible potentiometer

Specifications

Measuring Ranges

  • O₂ ppm Range: 0-100 ppm
  • 0-1,000 ppm
  • 0-10,000 ppm
  • O₂ % Range: 0-1 %, 0-25 %
  • Range must be specified at time of ordering

Calibration

  • With calibrated gas, air

Accuracy

  • ± 2 % FSD T = constant
  • ± 5 % FSD 0 > T > 50° C

Resolution

  • 0.1 ppm at 0-100 ppm
  • 1 ppm at 0-1,000, 0-10,000 ppm
  • 0.1%

Response Time

  • 90 % FSD at 25° C
  • 0-10 ppm < 45 s
  • 0-100 ppm < 20 s
  • 0-1,000 ppm < 10 s

Operating Temperature

  • 0° to 50° C

Pressure

  • 0.1 - 1 bar

Signal Output

  • 4-20 mA/dc

Alarm Value

  • 3.8 mA (Standard)
  • 23 mA (Optional)

Voltage

  • 10 - 35 Vdc

Load Range

  • 100 to 750 ohm
  • Typical 470 ohm

Display

  • 6 Digits, alphanumeric with bar graph

Oxygen Sensor

  • Micro-Fuel Cell

Housing

  • IP65

Dimensions

  • 120 mm x 160 mm x 65 mm (L x H x W)
  • 4.7” x 6.3” x 2.5”

Weight

  • 1.2 kg (206 lb)

Options

  • Different sensors available
  • 5 different ranges
  • 316 Stainless Steel cell Bl

Model Types

  • OxyT: Standard
  • OxyT-H: Works with the Hart Protocol
  • OxyT-RA: Right angle display
  • OxyT-SS: Panel mount sampling system, Panel size: 38.1 cm (15”) square, Weight: 3.2 kg (7 lb)
  • OxyT-PSS: Portable sampling system with power supply and pump option, Weight: 6.8 kg (15 lbs)
  • OxyTrans-EX: Explosion Proof, 2-Wire loop powered transmitter (ATEX approved)

  • Air Separators