Yodify Product Library
Instantly add the RPM10 PresSura Hospital Room Pressure Monitor and 50,000+ more product series to your store or catalog
Book Your Demo and See How
or create your store
RPM10 PresSura Hospital Room Pressure Monitor
Brand: TSI IncorporatedThe PresSura™ Model RPM10 Room Pressure Monitor is designed to accurately measure pressure differentials in hospital isolation rooms and operating rooms. Backed by TSI's unique thermal-based pressure sensor, the PresSura products offer the most accurate, precise, and stable measurement available.
These products continuously measure room pressure, verifying proper airflow and pressure direction for both Airborne Infection Isolation (AII) Rooms and Protective Environment (PE) Rooms. PresSura Products help hospitals ensure safety for patients and staff.
A local monitor provides easy visual indication of room status. If pressure is lost, the Model RPM10 activates audible and visual alarms, alerting staff to the alarm condition. The RPM10 features adjustable alarm delays to avoid nuisance alarms. When the room is unoccupied, the RPM10 can be changed to no isolation mode, deactivating all alarms.
Features- Stand-alone room pressure monitor verifies HVAC system operation
- Audible and visual alarms warn staff of unsafe conditions
- Complies with ANSI/ASHRAE/ASHE Standard 170-2013, Ventilation of Health Care Facilities
- Meets the Facility Guidelines Institute (FGI), Guidelines for Design and Construction of Health Care Facilities
- Meets US Centers for Disease Control (CDC), Guidelines for Environmental Infection Control in Health Care Facilities
- Meets United States Pharmacopeia (USP), General Chapter , Guidebook to Pharmaceutical Compounding - Sterile Preparations
- Accurate pressure measurement based on TSI's proven thermal anemometry
- Easy-to-read color touch screen display provides simple visual indication of room status
- Seamlessly integrates into building automation systems
- Accepts keyswitch to change room mode to neutral/no isolation
- Easy configuration via local programming on display
- Passwords prevent unauthorized access to monitor functions
- Compatible for use with PresSura NSD8 Nurses Station Monitor for easy central monitoring of all of the pressure sensors in an area
Specifications
Brand
Communication
- BACnet MS/TP
- LonWorks®
- Modbus
Range
- Bi-Directional Through-the-Wall Sensor: -0.20000 to +0.20000 in H₂O (-49.8 to +49.8 Pa)
Accuracy
- ± 10 % of Reading ± 0.00001 in H₂O (± 0.0025 Pa)
Resolution
- 5 % of Reading
Operating Temperature
- 32° to 120° F (0° to 49° C)
Temperature Compensation Range
- 55° to 95° F (13° to 35° C)
Alarm
- High Alarm: -0.19500 to +0.19500 in H₂O (-48.6 to +48.6 Pa)
- Low Alarm: -0.19500 to +0.19500 in H₂O (-48.6 to +48.6 Pa)
Display
- 4.3 in. (10.9 cm) Diagonal, Resolution: QVGA
Input
- Room Occupancy: Open/Closed Signal
- Door Switch: Open/Closed Signal
- Room Pressure Sensor: 0-10 Vdc
- Pressure Transducer: 0-10 Vdc
- Remote Keyswitch: Open/Closed Signal
- Exhaust Flow: 0-10 Vdc
- Supply Flow: 0-10 Vdc
Output
- Contact Outputs: Low Alarm and High Alarm / Room Mode, SPST (NO), Max. Current 2 A, Max. Voltage 220 Vdc/250 Vac
- Optional Analog Output: Flow or Room Pressure Differential 0-10 Vdc or 4-20 mA
Power Input
- 24 Vac, 50/60 Hz, 15-40 Vdc, 5 Watt Max.
Dimensions
- 7 in. x 5.875 in. x 1.75 in. (17.8 cm x 14.9 cm x 4.4 cm)
- Protrusion from Wall 0.625 in. (1.6 cm)
- Mounts to Triple-gang Standard Depth Electrical Box
Certifications & Approvals
- UL
- Airborne Infection Isolation (AII) Rooms - Negative Pressure
- Protective Environment (PE) Rooms - Positive Pressure
- Compounding Pharmacies
- Pandemic Preparedness Rooms
- Intensive Care Units
- Laboratories and Vivariums
- Burn Units
- Laundry Areas, Food Prep, Construction