The AMI Model 60 is the ideal solution for measuring percent level oxygen in in a general-purpose environment. 

In a very small size, and at very low cost, it provides a basic electronic package with a single range voltage output using an AMI electrochemical sensor. 

The model 60 is not affected by flammable gases and it may be used with practically any background gas. 

It is temperature compensated over its operating range and it may be used outdoors since its enclosure and connections are weather-proof. 

It needs between 7-24 Vdc power at less than 20 mA. 

Its sensor is unaffected by flow rate changes over the range of 0.1 to 5 SCFH.
Features
  • Percent oxygen in flammable and non-flammable samples
  • General purpose self-contained oxygen sensor
  • Patented AMI sensor
  • Temperature compensated
  • Virtually unaffected by humidity changes
  • Virtually unaffected by flow rate changes
  • Calibration interval 6 months
  • Very small size
  • Standard 0-2.5 V output
  • Very low cost

Specifications

Operating Temperature

  • 0° to 43° C (32° to 110° F), Wet Gas

Power Requirement

  • 7-28 Vdc (Nominally 24 Vdc) Typically 20 mA at 12 V with No External Draw from the 5 V Suppl.

Range

  • 0 to 1 %
  • 0 to 100 %
  • 0 to 25 %
  • 0 to 50 %

Repeatability

  • ± 1 % of Full Scale at Constant Temperature

Sensitivity

  • 0.5 % of Full Scale