Yodify Product Library
Instantly add the Diagnostic Thermocouple & mA Calibrator PIE 422Plus and 50,000+ more product series to your store or catalog
Book Your Demo and See How
or create your store
Diagnostic Thermocouple & mA Calibrator PIE 422Plus
Brand: Practical Instrument Electronics, IncCalibrate all your thermocouple instruments AND your milliamp loops with the PIE 422Plus. It has all the functions of a 14 type thermocouple calibrator and can source 0 to 24 milliamps or simulate a two wire transmitter like a milliamp calibrator.
You may also perform stand alone calibrations of two wire transmitters by simultaneously supplying the thermocouple input signal while supplying the 24 volts to power up the transmitter and monitoring the milliamp output signal.
Powers HART™ transmitters with built-in 250 ohm resistor simplifying hookups with HART communicators.
Features- Two Calibrators in One
- Troubleshoot loop problems
- Calibrate directly in temperature (°C & °F)
- Calibration Lab Accurate & Stable
- Verify Heat Treating Uniformity Survey Recorders and perform System Accuracy Tests
- Checkout all your 4 to 20 mA loops & instruments
- Easy to use in the Plant, Field or Shop
- Fast calibration with automatic output stepping
Specifications
Operating Temperature Range
- -25° to 60° C (-10° to 140° F)
Relative Humidity Range
- 10 % ≤ RH ≤ 90 % (0 to 35° C), Non-Condensing
- 10 % ≤ RH≤ 70 % (35 to 60° C), Non-Condensing
Temperature Drift
- ± 0.01 % of Span outside of 23° C ± 10° C (73° C ± 18° F)
Size
- 5.63 x 3.00 x 1.60 in, 143 x 76 x 41 mm (L x W x H)
Weight
- 12.1 ounces, 0.34 kg (including boot & batteries)
Batteries
- Four AA Alkaline 1.5 V (LR6)
Battery Life
- 25 Hours Thermocouple, 8 Hours milliamp
Isolation: Voltage
- 60 V RMS between all milliamp functions/Read Vdc and Source Vdc/Thermocouple
Normal Mode Rejection
- 50/60 Hz, 50 dB
Common Mode Rejection
- 50/60 Hz, 120 dB
Optional NiMh Rechargeable Battery Kit
- Charger, four NiMh batteries
Low Battery
- Low battery indication with nominal 1 hour of operation left
Protection Against misconnection
- Over-voltage protection to 60 V dc (rated for 30 seconds)
Display
- High contrast graphic liquid crystal display. LED backlighting for use in low lit areas.
Ranges and Resolution
- 0.000 to 24.000 mA or -25.00 to 125.00 % of 4-20 mA
Accuracy
- ≤ ± (0.02 % of Reading + 0.003 mA)
Voltage Burden
- ≤ 2 V at 24 mA
Overload/Current Limit Protection
- 25 mA nominal