Single-Use Sensor Monitors
Brand: SciLog, IncMonitors for use with SciLog® Single-Use Sensors are available for the full range of sensors: SciPres® Pressure Sensors, SciCon® Conductivity Sensors, and SciTemp® Temperature Sensors.
In addition to displaying the outputs for its specific sensor type, each monitor will read and display the unique sensor serial number and factory calibration information recorded on each SciLog® Sensor's memory chip.
When connected to a PC using SciDoc™ Software, available separately, real-time documentation of sensor data can be completely automated.
SciPres® Pressure Sensor Monitors can handle up to 3 pressure sensor inputs (P1, P2 and P3) simultaneously and have a user selectable readout permitting the display of transmembrane pressure (TMP) and differential pressure (dP).
SciCon® Conductivity Sensor Monitors display conductivity and temperature for a single sensor, and display a conductivity range of 1µS/cm to 200mS/cm and a temperature range of 0° to 50°C.
SciTemp® Temperature Sensor Monitors can read up to 2 sensor inputs simultaneously and displays T1, T2 and differential temperature (DT) with an accuracy of +/- 0.5° C in the 4.00° to 70.00° C range.
3 TTL alarm switches on the SciTemp® Monitor and 4 TLL alarm switches on the SciPres® and SciCon® Monitors, with relevant user selectable high/low limits for each monitor.
Monitors have both analogue and digital (RS-232) outputs.
1 - 3 of 3 Part Numbers
1 - 3 of 3 Part Numbers
Specifications
Brand
Industry
- Life Sciences
Height
- 2.75”
- 69.85 mm
Weight
- SciCon: 0.56 kg, 1.23459 lbs
- SciTemp: 0.59 kg, 1.30073 lbs
- SciPres: 0.62 kg, 1.36687 lbs
Depth
- 5.75”
- 146.05 mm
Width
- 6”
- 152.4 mm
- Growth Media Preparation
- Bioreactor Maintenance
- Cell Culture Harvest
- Buffer Preparation
- Product Filtration
- Tangential Flow Filtration
- Virus Control
- Bulk Drug Dispense
Documentation
Literature and Reference Materials
- Biopharmaceutical Industry Overview Guide --GL BP 05 01/20 Rev 2A pdf 1.2 MB
- SciLog SciCon Installation, Operation & Maintenance Instructions--IOMI-AK1-002 pdf 4.6 MB
- SciLog SciPres Installation, Operation & Maintenance Instructions--IOMI-AK1-001 pdf 848 KB
- SciLog SciTemp Installation, Operation, & Maintenance Instructions--IOMI-AK1-003 pdf 1.7 MB