Yodify Product Library
Add the Ammonia Analyzer - Economical to your store or catalog
Book Your Demo and See How
or create your store
Ammonia Analyzer - Economical
Brand: Los Gatos Research, IncLos Gatos Research's (LGR's) Ammonia Analyzer provides sensitive measurements of ammonia in ambient air or in process flows containing ethylene with extremely high precision and sensitivity. No longer do you have to spend a lot of money or wait a long time or rely on scrubbers to measure ammonia with high sensitivity – LGR’s Ammonia Analyzer provides direct measurements every second with ppb-level precision. In addition, the Ammonia Analyzer can report measurements quickly (fast response time) over a very wide range of concentrations. The instrument, based on LGR’s cavity enhanced laser absorption spectroscopy, is simple to use, inexpensive to operate, and includes all components (internal vacuum pump, keyboard, mouse, video monitor) to start recording data within minutes.
LGR’s Ammonia Analyzer provides the high quality data necessary for the most demanding applications including semiconductor process monitoring and atmospheric trace gas monitoring. Moreover, only LGR’s analyzers provide reliable measurements (with guaranteed specs) at extremely high concentrations (to 10 ppm ammonia in air). In addition, with optional ethylene capability, the instrument can report measurements of ammonia directly in 'pure' ethylene flows (without the need for any scrubber or background subtraction).
Features- 1 Hz measurements: allow observation of transient and time varying flows
- Accurate measurements over a wide range of concentrations
- High resolution absorption spectra viewable for diagnostics
- Reports ammonia and water vapor simultaneously
- Reports water vapor up to 60,000 ppm
- Low power: ideal for field work
- Enhanced performance model provides lowest drift and highest accuracy
- Reports ammonia in ethylene flows (optional)
Specifications
Brand
Measurement Range
- H₂O: 40 to 60,000 ppm, Non-Condensing
- NH₃: 1 to 10,000 ppb
Communication
- Analog
- Ethernet
- RS-232 (Digital)
- USB
Mounting
- Rack-Mounted
Response Time
- < 2 seconds (With Optional External N920 Vaccum Pump)
- 10 seconds (with Standard Internal Pump)
Humidity
- Ambient: 0 to 100 % RH, Non-Condensing
Operational Range
- H₂O: 0 to 70,000 ppm, Non-Condensing
- NH₃: 0 to 200 ppm
Precision (1σ, 1 Sec / 10 Sec / 100 Sec)
- 0.6 ppb / < 0.2 ppb / < 0.12 ppb
Power Requirement
- 115/230 Vac, 50/60 Hz
Power Consumption
- 100 W - Standard Model
- 300 W - Enhanced Performance (EP)
Operating Temperature
- 0° to 50° C (32° to 122° F)
- 5° to 45° C (41° to 113° F)
Dimensions
- Rackmount (Standard model): 8.75" x 19" x 24"
- Rackmount (Enhanced Performance model): 14" x 19" x 24"
Weight
- 65 pounds (Standard models)
- 90 pounds (Enhanced Performance model)