Yodify Product Library
Instantly add the piD-TECH eVx Photoionization Sensors and 50,000+ more product series to your store or catalog
Book Your Demo and See How
or create your store
piD-TECH eVx Photoionization Sensors
Brand: MOCONAs one of the most widely used gas detection techniques, the Photoionization Detector (PID), is commonly applied in both portable and fixed instruments for detection of a wide variety of organic compounds and some inorganic gases in ambient air.
MOCON’s patented, award-winning piD-TECH® eVx™ plug-in photoionization sensor provides complete photoionization detection capabilities and is designed to be mechanically similar to major brands of electrochemical sensors.
Outstanding features make the piD-TECH line of sensors the right choice for OEM manufacturers looking to include VOC detection capabilities in their handheld, mobile, or fixed devices.
Offering better value and design flexibility for OEM markets and incorporating state of the art technology that cannot be matched in the marketplace, the piD-TECH eVx detection capabilities and minimum detection quantity (MDQ) come in five ranges depending upon application requirements.
Our dedicated sensor engineers will assist you in integrating our intrinsically safe photoionization sensor into your products, reducing the time and cost associated with product development.
Features- Intrinsically safe: UL, CAN/CSA, ATEX, IECEx Certified
- Integral shielding
- Lamp re-ignition circuit for fast, repeatable start-up
- Reliable long lamp life: >6000 hours
- Internal input voltage regulation for greater signal stability
- Dual filter protects from aerosols and particulates
- Easily cleaned and field serviced
- OEM integration support available
Specifications
Brand
Sensor Ranges
- Range: 2 - 10,000 ppm
- Detection Limit: 0.5 ppb - 500 ppb
- T90 Response: < 2 sec or < 4 sec
Operating Specifications
- Temperature Range: -20° C to 60° C (-4° F to 140° F)
- Relative Humidity Range: 0 to 90 % Non-Condensing
- Humidity Response: < 1 % @ 90 % RH
- Humidity Quenching Effect: < 15 % @ 90 % RH
- Accuracy: ± 3 % of Reading, with Constant Temperature and Pressure
Electrical Characteristics
- Supply Voltage: 3.2 V to 5.5 V (Input Voltage Regulator Included)
- Current: 24 mA to 38 mA
- Power Consumption: 80 mW to 200 mW (Dependent upon Supply Voltage)
- Output Signal: 0.045 V to 2.5 V Linear
Physical Characteristics
- Weight: < 8 grams
- Dimensions: 20.4 mm Dia. x 17.6 mm H (pins 4.78 mm H)
- Serviceable Parts: Lamp, Detector Cell, Filters, Cap, Spacer
- Lamp Life: Guaranteed 6000 hours
- Onboard Filters: Removes Liquids and Particles
- Warranty Period: 2 years, not Including Consumables
Certifications & Approvals
- USA: UL 913, 8th Edition. Intrinsically Safe Apparatus and Associated Apparatus for use in Class I, II, and III, Div. 1, Hazardous (Classified) Locations
- Europe: ATEX directives: EN 60079-0:2018, EN 60079-11:2012, (EX) II 1 G Ex ia IIC Ga, -20° C - 60° C, ATEX certificates: DEMKO 13 ATEX 1304446U Rev. 6, C E 0539
- Other: IECEx Standards: 60079-0 Ed. 7, 60079- 11 Ed. 6, Certificate IECEx UL 13.0050U Issue: 6, CAN/CSA C22.2 No.157-92
- Patents: US Pat 6,646,444 Japan Pat 3,793,757
- Environmental Monitoring
- AMC in Cleanrooms
- Air Quality Monitoring
- Fugitive Emission Monitoring
- Toxic Gas Monitoring
- Indoor Air Quality
- Leak Detection
- OEM PID Sensors