Yodify Product Library
Instantly add the Baseline 9000 Total Hydrocarbon Analyzer and 50,000+ more product series to your store or catalog  
Book Your Demo and See How
or create your store

Baseline 9000 Total Hydrocarbon Analyzer

Brand: MOCON

The Baseline® 9000 Total Hydrocarbon Analyzer is a microprocessor-based instrument designed for continuous ambient or process hydrocarbon gas measurement in environmental or industrial settings.

The total hydrocarbon analyzer can be purchased in a variety of configurations with internal components for single or multi-point sampling (with or without a sample pump) for pre-filtered non-condensing samples.

Detection limits down to < 10 ppb. User-programmable ranges from 1 ppm up to 100% as methane (CH₄) are factory-configured per the customer’s application to facilitate installation and setup.

Using a Flame Ionization Detector (FID), AMETEK MOCON's FlowGuard electronic control regulates the delivery of fuel, air, and a small portion of the sample gas, to the FID.

During the combustion process, organic or hydrocarbon-based gases in the sample are ionized, detected by the instrument, and then reported as a concentration.

The automatic calibration feature enhances the long-term analytical stability of the instrument. All instrument parameters are reported clearly and continually refreshed on a large, graphical LCD display.

Using analog, digital, and logic output communication capabilities, analytical information from the analyzer can be acquired using an external PC and a simple communications program such as Windows® HyperTerminal or the analyzer can output binary or ASCII formats directly to a data acquisition system or PLC.

Every Baseline 9000 analyzer includes AMETEK MOCON's free PC utility 9000 Keeper used for storing and uploading multiple methods, as well as sending configuration settings, directly to the analyzer.
Features
  • Automated Control
    • Automatic FID ignition
    • FlowGuard control of fuel, air, and sample with automatic shut-off
    • Electronic back-pressure regulator with sample bypass system
    • Automatic calibration at user-defined intervals
    • Accurate, continuous, detection
  • Versatile Platform
    • Graphic LCD display with easy to use menu
    • Benchtop or rack-mountable
    • Single or multi-point sampling
    • Ethernet and serial customizable output
    • Programmable analog output ranges
    • Programmable relays for diagnostics, concentration, alarms, and events

Specifications

Brand

Detector

  • Flame Ionization (FID)

Ranges

  • User Definable Based Upon Calibration (as Methane):
  • Very Low - 0.01 ppm to 200 ppm
  • Low - 0.1 ppm to 2,000 ppm
  • Medium - 0.3 ppm to 20,000 ppm
  • High - 0.003 % to 100 %

Accuracy, Repeatability

  • ± 1 % Full Scale

Drift, Zero

  • ± 0.025 % Full Scale, 24 hours

Drift, Span

  • ± 1 % Full Scale, 24 hours

Response Time

  • T90 < 5 seconds

Sampling

  • Internal Single Point or Multi-Point, Pump Optional

Alarms

  • Multilevel Concentration and Fault Alarms, Audible and Visual

Calibration

  • Programmable Automatic or Manual Calibration

Support Gas Requirement

  • UHP Hydrogen (H₂) - 35 cc/min
  • Zero Air - 175 cc/min
  • Span Gas - Methane is Typical
  • Fuel Blend Options Available

Display

  • Graphical LCD, 3.4” x 4.5” (8.64 cm x 11.43 cm)

Outputs, Standard

  • RS-232
  • LAN/Ethernet
  • Analog, 4 Programmable 0-20 mA/4-20 mA Isolated Output
  • Relays, 5 Programmable Form A Relays Rated to 3A @ 230 Vac

Outputs, Optional

  • 3 Additional Analog Outputs
  • 5 Additional Relays

Operating Temperature

  • 32° to 104° F (0° to 40° C)

Operating Humidity

  • 0 to 95 % (Non-Condensing)

Configuration

  • Benchtop or 19” (48.3 cm) Rack-Mount, 3U

Accessories

  • Preventive Maintenance Kit - Includes all parts potentially needed on-hand
  • Gas Generators - Hydrogen generators, Zero Air generators
  • Software - 9000 Keeper provided for method switching
  • Service & Support Options - Start-up and training, 1 day onsite

Dimension & Weight

  • 5.25” H x 19” W x 16.25” D (13.02 cm x 48.26 cm x 41.28 cm), < 20 lbs (9.07 kg)

Electrical Requirements

  • Voltage: 100-240 Vac 50/60 Hz, 1 A

  • Environmental Monitoring
    • AMC in Cleanrooms
    • Air Quality Monitoring
    • Fugitive Emission Monitoring
  • Toxic Gas Monitoring
    • Indoor Air Quality
    • Leak Detection
    • OEM PID Sensors
  • Industrial Process Analysis & Control
    • Beverage Gas Monitoring
    • Industrial Gas Mixing Control
    • Process Gas Analysis
    • Specialty and Industrial Gas Monitoring
    • Surface Logging Analysis
Instantly populate your store or catalog 