ASM Series Process Multi-Scanner
Brand: JOFRA CalibrationWe offer a signal multi-scanner which provides a time-saving, automatic solution to calibrate multiple temperature or pressure sensors simultaneously.
The scanner may be combined with our dry-block or liquid temperature calibrators, DTI temperature indicators, pressure calibrators, and ASC multifunction process calibrator.
Use any place where temperature and/or pressure measurement is critical and there is a need to create traceable documents for sensor calibration.
The Advanced Process Multi-Scanner offers unique time-saving and automation solutions to calibrate multiple pressure or temperature sensors simultaneously.
Each ASM contains an eight channel scanner controlled by Jofracal software on a PC. It can handle signals from 2-, 3-, and 4- wire RTDs, TCs, transmitters, thermistors, and temperature and voltage switches.
Up to 3 ASM units can be stacked to calibrate up to 24 sensors at the same time.
The ASM may be combined with any JofraCal compatible pressure calibrator, dry-block or liquid calibrators, DTI reference thermometers, and ASC multi-function process calibrators.
Use any place where temperature measurement is critical and there is a need to create traceable documents for sensor calibration.
Features- Automatic procedures reduce human error
- Dedicate one input channel to a reference sensor with accuracy to 0.026°C
- Uses True Ohm technology to eliminate EMF from cables, sockets, and sensors
- A models use a separate measuring unit as the standard, and B models have built-in measurement capabilities
- Combined with JofraCal or JofraLog you can:
- Set-up, execute, print, and save traceable calibration data
- Calibrate up to 24 sensors using a dry-block calibrator as the reference
- Calibrate using a high accuracy temperature reference probe directly with the ASM
- Data-log up to 24 sensors simultaneously
- Inputs for Thermocouple Calibration
Specifications
Brand
Power Supply
- External AC/DC Adapter
Input
- 90-254 V, 45-65 Hz
Output
- 30 V ± 2 % Regulated DC, Max. 30 W
Scanning Rate
- Maximum 5 Channels per second
Weight & Instrument Size
- Weight: 2.3 kg/5.07 lb
- (LxWxH): 250 x 249 x 69 mm / 9.8 x 9.8 x 2.7 in
Shipping (with Carrying Case)
- Weight: 6.3 kg/13.9 lb
- (LxWxH): 350 x 560x180 mm / 13.8 x 22.1 x 7.1 in
Shipping (without Carrying Case)
- Weight: 4.4 kg/9.7 lb
- (LxWxH): 350 x 560x180 mm / 13.8 x 22.1 x 7.1 in
Miscellaneous
- Serial Data Interface: RS-232
- Specification Temperature: 20 to 26° C (68 to 79° F)
- Operating (Ambient) Temperature: 0 to 40° C (32 to 104° F)
- Storage (Ambient) Temperature: -20 to 50° C (-4 to 122° F)
- Humidity: 0 to 90 % Relative Humidity @ 30° C
- CE Conformity: EN 61326
Transmitter Supply
- Output Voltage: 24 Vdc +10 %
- Output Current: Max. 28 mA
- 8 individual current sources. Permanent on, no switching
Accuracy Automatic Cold Junction Compensation
- ASM-801/802 - ± 0.20° C (± 0.36° F): 20 to 26° C (68 to 79° F) @ Ambient Temperature
- ASM-803 - ± 0.50° C (± 0.90° F): 20 to 26° C (68 to 79° F) @ Ambient Temperature
- Temperature Drift Outside 20 to 26° C. 0.05° C/°C 0.05° F/°F)
Input Specifications - A Models
- A-Models When Used with Other Equipment - Accuracies from the connected instruments has to be added
- RTD 4-Wire: 2.5 ppm reading (0-400 ohm), 15 ppm reading (400-4000 ohm)
- RTD 3-Wire: 2.5 ppm reading + 50 mohm (0-400 ohm), 15 ppm reading + 50 mohm (400-4000 ohm)
- mA: 1 ppm reading (0-24 mA)
- MV, V: 2uV
Transmitter Input mA (B Model)
- Range: 0 to 24 mA
- Accuracy (12 months): ± 0.01 % Reading. +0.01 % Full Scale
Voltage Input VDC (B Model)
- Range: 0 to 12 Vdc
- Accuracy (12 months): ± 0.005 % Reading. +0.01 % Full Scale
Switch Input
- Switch Dry Contacts
- Test Voltage: Max. 2.5 Vdc
- Test Current: Max. 0.8 mA
RTD Input Specifications
- Signal Type: 2-, 3-, 4-Wire true ohm RTD input
- Signal Range: 0-400 Ω (PT10/PT50/PT100)
- Accuracy (12 months): ± 0.002 % Reading. +0.002 % Full Scale
- Signal Range: 0-4000 Ω (PT200/PT500/PT1000)
- Accuracy (12 months): ± 0.002 % Reading. +0.005 % Full Scale
- For 3-Wire input add 50 mΩ assuming all three RTD leads are matched. For 2-Wire add 100 mΩ
Thermocouple Specifications
- Signal Range: -10 mV - 78 mV
- Accuracy: ± (0.005 % of Reading. + 0.005 % of Full Scale)