Yodify Product Library
Add the MT OM Pressure Transmitter with Local Readout, for Homogenizer DS 4" (100 mm) to your store or catalog
Book Your Demo and See How
or create your store
MT OM Pressure Transmitter with Local Readout, for Homogenizer DS 4" (100 mm)
Brand: Nuova FimaThese instruments are manufactured to monitor the electrical operations on hermetically sealed systems containing Sulphur Hexafluoride gas (SF6).
The indication and the electrical operations are calibrated to the gas density (isochore) according to the relation pressure-temperature.
The M5 model is suitable for indoor installation while the M6 model has been designed for the outdoor installation as well.
The oil filled executions are particularly suitable for installation when vibrations are apparent
Specifications
Brand
Accuracy (% VFS)
- Local Readout, ≤ 1.0 (≤ 1.6 for Pressure Ranges > 8700 psi - 600 bar)
- Transmitter, ≤ 0.5
Ambient Temperature Range
- 14° to +149° F (-10° to +65° C)
Annual Drift
- ≤ 0.2 % of Span
Bourdon Tube
- Standard Model: AISI 316L Stainless Steel. Seamless Tube
Compensated Temperature Range
- 14° to +176° F (-10° to +80° C)
Designation Code
- Standard Model: S1 as Per EN 837-2
Dial
- Standard Model: Aluminum, White with Black Markings
Diaphragm
- Standard Model: AISI 316L Stainless Steel.
Diaphragm Seal
- Standard Model: AISI 316L Stainless Steel. with Finishing Ra ≤ 0.8 µm (Welded Parts Included)
Electric Connection
- Standard Model: Junction Box as Per VDE with Exit for Cables Ø 0.27 to 0.51” (7 to 13 mm)
Filling Liquid
- Filled Model: Dielectric Oil
IP Protection Rating
- Filled Model: IP67 as Per EN 60529 / IEC 529
- Standard Model: IP55 as Per EN 60529 / IEC 529
Movement
- Standard Model: Stainless Steel
Other Features
- Filled Model: as Standard Model
Output Signal
- For Pressure Ranges > 8700 psi (600 bar): 4-20 mA
- For Pressure Ranges ≤ 8700 psi (600 bar): 4-20 mA, 0 to 5 Vdc, 0-10 Vdc
Over Pressure
- Not Suitable
Pointer
- Standard Model: Adjustable, Aluminum, Black
Process Temperature
- 14° to +248° F (-10° to +120° C)
- Maximum
Range
- From 0 to 1500 to 0 to 20,000 psi, Relative (From 0 to 100 to 0 to 1600 bar or Equivalent Units)
Ring
- Standard Model: Stainless Steel, Bayonet Lock
Sensor
- Standard Model: Ceramic Thick Film or Stainless Steel Thin Film
Sensor Calibration
- Limit-Point as Per DIN 16086
Span Calibration
- ± 10 % of Span Typical
Thermal Drift
- ≤ 0.011 % of Span/°F (≤ 0.02 % of Span/°C)
Window Material
- Standard Model: Tempered Glass
Working Pressure
- 75 %, Max.
Zero Calibration
- ± 10 % of Span Typical